14
Amico Mobility Solutions Corporation
Emergency Lowering
In the event that the DOWN button on the hand control does not function, or in power failure situations, the patient
may be lowered by pressing and holding down the RED emergency stop button located on the top of the lift.
Continue to press down until the patient is safely lowered to the desired position. The lift will beep as you continue
to press down on the red button and will continue beeping until the button is released after the desired lowering has
been achieved.
NOTE: The emergency lowering button does not provide a raising function. The failure of any of the lowering devices
should be reported to Amico Mobility.
!
In an event of emergency when normal lowering system of the lift malfunctions and the “emergency
lowering” function is used, the lift must be reset by a qualified lift technician before re-use.
Manual Emergency Lowering
The manual emergency and raising should only be used if the emergency lowering does not work. A proper safety
ladder or stool may be required to remove the plug from the cover. Remove the round plug from the lift cover and use
3/16" or 2.5 mm Allen key to rotate the motor in the up or down direction.
!
Do not attempt to use the lift while using manual lowering.
Remove the round plug from the lift cover and use 3/16" or 2.5 mm Allen key to rotate the motor in the up or down
direction.
Overspeed Cam
The Overspeed Cam brake is made of a metal bar fixed to the drum. In case of gear or motor breakage, the centrifugal
force created will block the bar against the frame.
Emergency RED button
Continue to press down until the
patient is safely lowered to the
desired position.