INTRODUCTION

The Amico medical gas alarm system (ALERT-1) incorporates the latest digital technology in alarm

arid surveillance systems.

The system has been designed primarily for user flexibility and reliability.

• Digital read-out for every gas service measured.
•  Bar-graph display for trend indication.
•  HI/LOW alarm indication with adjustable repeat times.
•  Interface to building management computer.
•  Easily expandable for additional gas services.
•  

NO 

moving parts    v e r y  reliable.

• Modular construction    e a s e  of maintenance.
•  Plexiglass front covere a s y  to clean.
•  Adjustable repeat alarm

Optional computer interface.

• Meets requirements of CSA 2305.1.

ALERT-1 is divided into four main modules which plug
into the alarm box.

1. POWER SUPPLY MODULE

This 'module converts the 115 VAC

supply voltage into 24 VDC required for

the alarm system and contains the

audible alarm and lamp test circuitry.

2. PRESSURE/VACUUM

DISPLAY MODULE

This module provides a digital display
of the actual pressure or vacuum of a
gas being monitored. In addition a gas

movement trend and HI/LOW alarm
section is provided for quick identifica-

tion.

3. SENSOR MODULE

This module contains the transducer
which converts the pressure (vacuum)
source into an electric signal that is
displayed  b y   t h e  pressure display

module.

=

I

 -.I-GAS

Features include:

MEDICAL  C A S

ALARM SYSTEM

(((co))

P0v,'ER  L I G H T

AUDIBLE A L A R M

...4.1,!2  T E S T   &

  I = "  I   - - I - A L ; - R m  SILENCE

ANDO  I N S T R U + 0 0 5

14101070. 

CANADA

POWER MODULE

—.1---PRESSURE

& TREND

I N D I C AT O R

LDV  I

GAS

!]—•--VACUUM

A L A R M   &   T R E N D

NCIIInAL

I N D I C AT O R

PRESSURE DISPLAY 

V

A

C

U

U

M

 DiSPL

Содержание ALERT-1

Страница 1: ...ACorporation M I C O ALERT 1 DIGITALALARM SURVEILLANCE SYSTEMS FORTHE HEALTH CARE INDUSTRY SUITABLE FOR Hospitals Laboratories Nursing Homes Clinics INSTALLATION MAINTENANCE MANUAL...

Страница 2: ...module 5 3 Sensor Module 5 4 Status Module 6 5 Computer Interface Module 6 Electrical Connections 7 Trouble Shooting 8 Drawings 9 Wiring Diagrams Area Alarm 10 Master Alarm Computer Interface 1I Powe...

Страница 3: ...ules which plug into the alarm box 1 POWER SUPPLY MODULE This module converts the 115 VAC supply voltage into 24 VDC required for the alarm system and contains the audible alarm and lamp test circuitr...

Страница 4: ...N O R R O W E MASTER STATUS MODULE ALERT 1 is factory calibrated for Normally Closed Circuit NC operation so that the alarm function will be initiated on an open circuit In this mode the alarm system...

Страница 5: ...wall remove the dust cover from the front Using silver brazing techniques con nect each pigtail to its appropriate gas Vacuum while ensuring that the bot tom of the pigtail is wrapped with a damp clo...

Страница 6: ...tor c Flip the power module so that it hangs by its cable just below the box The circuit with all the components should now be visible d Bring the power lines which protrude through the transformer co...

Страница 7: ...le in the box in the wall The connector should click into place Fig 4 b Remove the display module from its box and plug the connector atthe end of the tail into the appropriate connector onthe motherb...

Страница 8: ...minal blocks on strip J4 as per the wiring diagram shown on page 11 4 COMPUTER INTERFACE MODULE The computer interface module is a piggyback board that fits on top of the status module and plugs into...

Страница 9: ...plied the display lights up and the buzzer should sound when the push to test button is depressed For problem identification see the section on trouble shooting When each module has been checked indiv...

Страница 10: ...Ensure the gas going into the pigtails is being measured accu rately b Ensure the sensor board is connected to the display board at terminal J2 Fig 4 c Ensure the display board is plugged into the mo...

Страница 11: ...POSITION BOX I 3 3 0 0 4 1 I 3 3 0 0 3 250 11 125 A LIMO MEDICAL GAS ALARM STSILLI 0o 0110 10010 1 1100 10 11101110000010 0 N W 1 Q n LDV U N M rr _ 14 000 16 400 3 POSITION BOX 6 125 r _ 0 750 0 750...

Страница 12: ...WIRING DIAGRAM FOR AREA ALARM Power Module To Remote Alarm C Low High Back Box Mother Board Ribbon Cable Low High Ribbon Cable Pressure Module MP MOD L GAS Sensor Module 10...

Страница 13: ...3 POINT 2 NO 0 0 2 POINT I NO 0 0 1 C Field Equipment N C Common C N C C Common Common N C C Common Common N C C N C Common Common N C MP MOD E MAS Common Typical examples for closed circuit operatio...

Страница 14: ...DIAGRAM FOR 120VAC SUPPLY 120VAC Supply 24VDC to Manifold Fuse 1 2 Amp 0 0 N 0 0 0 cr N Back Box ransformer Power Module 12 24 Gt4D 15 MP MOD E POW Power Plug Rectifier Connection to Optional Remote...

Страница 15: ...SI 1 9 5 PSI Vacuum Low only 2 1 Hg 1 2 Hg 40 PSI 140 PSI TheAlarm is factory calibrated to the aboveset points The Alarm is also pre calibrated to ALTERNATE OPERATING set points for hospitals whosega...

Страница 16: ...ain the proper range You would then adjust R42 to obtain the desired values Calibration Procedure for the Pressure Display Module Setup Connect the display PCB to a well calibrated Sensor module in a...

Страница 17: ...g i t o n o f f select R22 D i s p l a y Zero a d j u s t m e n t R25 D i s p l a y Span a d j u s t m e n t R6 B a r g r a p h non standard gain a d j u s t m e n t R42 B a r g r a p h C e n t e r a...

Страница 18: ...y Corporation M I C O 8 30 Mural Street Richmond Hill Ontario L 4 B 185 CANADA Te l 905 764 0800 Fax 905 764 0862 Andersen Medical Gas 12 Place Lafi fitte Madisonville LA 70447 www TheMedicalGas com 1...

Отзывы: