69
Remote Interface Reference
SCPI Introduction
(see page 88 for more information)
PSwitch {0|1}
PSwitch?
VOLTage:PSwitch?
(see page 89 for more information)
QUench {0|1}
QUench?
(see page 90 for more information)
*ETE <
enable_value
>
*ETE?
*TRG
4.2 Programming Overview
The Model 420 conforms to the SCPI (Standard Commands for
Programmable Instruments) IEEE standard. The SCPI standard is an
ASCII-based specification designed to provide a consistent command
structure for instruments from various manufacturers.
The Model 420 also implements a status system for monitoring the state of
the Model 420 through the Standard Event and Status Byte registers.
4.2.1
SCPI Language Introduction
SCPI commands conform to a tree structure where commands are grouped
according to common keywords. For example, commands which set a
Model 420 setup or operating parameter begin with the keyword
CONFigure
. The keywords are shown in upper case and lower case to
indicate acceptable abbreviations. For the example keyword
CONFigure
,
the user may send either the abbreviated form of
CONF
, or the entire
keyword
CONFIGURE
. Any other form of the keyword is illegal and will
generate an error.
Switch Heater Commands and Queries
Quench State Control and Queries
Trigger Control and Queries
Содержание 420
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 12: ...x Rev 7 List of Tables ...
Страница 18: ...xvi Rev 7 Foreword Safety Summary ...
Страница 22: ...4 Rev 7 Introduction Front Panel Layout ...
Страница 24: ...6 Rev 7 Introduction Rear Panel Layout ...
Страница 30: ...12 Rev 7 Introduction Operating Characteristics ...
Страница 54: ...36 Rev 7 Installation Power Up Procedure ...
Страница 82: ...64 Rev 7 Operation Summary of Operational Limits ...
Страница 132: ...114 Rev 7 Appendix RS 232 422 Connector ...