background image

AMI Analyzer Manual 

Installation Procedure  

  10 

Operation 

Analyzer operation 

It is just as important that the analyzer’s sample system – the components used to 

control the gas flow – are perfect as the measurement methodology itself.  No matter how 

good the analyzer is, if you don’t connect the plumbing correctly you won’t get good 

results. 

Sample tubing and components 

Oxygen diffuses through plastics, to a greater or lesser extent.  Standard blue poly tubing 

will diffuse about 1ppm per foot into 1SCFH at 70°F.  Silicone tubing is very much worse.  

In general, use stainless steel tubing, or at least copper tubing, not plastic.  Use high 

quality compression fittings such as “Swagelock™” or “Gyrolock™” and high quality 

stainless steel filters, valves or regulators if necessary.  Make sure that they are 

assembled correctly and perform a thorough leak test on the sample system before use. 

The analyzer expects to see sample pressures between about 1 psig and 100psig.  Higher 

pressures will make the flow control valve very touchy.  If your pressure is higher than 

this use a regulator with a stainless steel diaphragm to bring the pressure down 

appropriately.  The analyzer is not very much affected by changes in flow rate and the 

internal flow meter is adequate for setting the flow.  Don’t use a high precision external 

flowmeter in front of the analyzer because oxygen will diffuse in through its O rings. 

The span gas pressure should also be held below 100psig.  If you permanently connect a 

span gas tank to it via a user-supplied three way valve, make very sure that the 

connection is leak tight as otherwise not only will the calibration be in error but also you 

will lose all of your span gas as it leaks out. 

Leaks 

If you suspect a leak due to high oxygen readings, change the flow rate and see how long 

it takes for the oxygen reading to change.  A higher flow rate will dilute the effects of a 

leak so that the reading will drop as you increase the flow (this is a very good way of 

making sure that you do not in fact have a leak, by the way).  Turn the flow up from 1 

SCFH to 2 SCFH using the front panel valve and see how long it is before the reading 

changes.  If it does so immediately, the leak is close to the analyzer.  If it takes longer, the 

leak is further away. 

Exhaust 

The exhaust port of the analyzer should be given at least a foot of tubing, but it should 

not be allowed to become pressurized, or the readings will be affected.  If the sample gas 

is innocuous, such as pure nitrogen, it can be vented into the room with no ill effects, if 

the room itself is ventilated.  Otherwise it should be vented outdoors or brought into a 

scavenging system.  If the latter, have the scavenging system suck in room air around the 

exhaust line so that the pressure at the exhaust is atmospheric.  A typical way of doing 

this is to have the scavenging system suck on a 1” pipe, and have the ¼” exhaust line 

from the analyzer extend six inches or so into the end of the pipe. 

Содержание 2001LC

Страница 1: ...AMI Oxygen Analyzer Manual Model 2001LC AMI Costa Mesa CA ...

Страница 2: ...ion and Operation 4 Receiving the analyzer 4 Installation 4 Location 4 Precaution 4 Installation Procedure 5 Connect gas and power lines 7 Interconnections 8 Alarm connections 8 Output connections 8 Serial connections 8 Sample Handling 9 Operation 10 Analyzer operation 10 Sample tubing and components 10 Leaks 10 Exhaust 10 Calibrating the analyzer 11 Output range concept 12 Front Panel Controls 12...

Страница 3: ...Bypass 14 Verify Span Factor 14 Read the Temperature 14 Alarm Functionality 14 Maintenance and troubleshooting 15 Maintenance 15 Periodic Calibration 15 Sensor Replacement 15 Sensor replacement cautions 16 Sensor replacement procedure 17 Bleeding a regulator 18 Troubleshooting 19 All oxygen applications 19 Specifications and Disclaimer 21 Specifications 21 Disclaimer 22 ...

Страница 4: ...the air contains 20 9 209 000ppm of oxygen and it can get into a pressurized pipeline through the smallest leaks Oxygen molecules will enter through a leak no matter the pressure or the nature of the gas in the line This analyzer is an exceptionally sensitive leak detector including those provided by improper installation Make sure you read this manual carefully prior to installation Caution Read ...

Страница 5: ...us area This manual covers software version 1 0 Features Compact size Unique patented cell block Auto ranging display with user selectable output range Front panel sensor access Optional air or span gas calibration no zero gases required Virtually unaffected by hydrocarbons or other oxidizable gases High accuracy and fast response Large liquid crystal display Backed by a two year warranty excludin...

Страница 6: ...h much faster come down times typically under twenty minutes to 10ppm from a 1 minute air exposure Sensor Warranty The sensor is warranted to operate for a period determined by its class If the sensor ceases to operate correctly before this time has elapsed contact AMI for a return authorization for evaluation If there is any evidence of defective material or workmanship the sensor will be replace...

Страница 7: ...panel in a general purpose area It should be mounted at a suitable viewing level Refer to the drawing figure 1 showing the analyzer dimensions It is not suitable for use in a hazardous area or with flammable gases Although the unit is RFI protected do not to mount it close to sources of electrical interference such as large transformers motor start contactors relays etc Also avoid subjecting it to...

Страница 8: ... 3 Deal with any potential condensation or liquid contamination issues 4 Connect the sample line to the sample inlet port with ss tubing 5 Pressurize the sample line to line pressure between 1psig and 20psig 6 Leak check every fitting and weld from the analyzer inlet to the sample tap 7 Connect vent line to outside or a suitable purge system 8 Connect power relay contacts and analog output If usin...

Страница 9: ... spray e Disconnect the sample tubing from the inlet port either by physically removing it or by using an external three way valve f Flow calibration gas WHILE you are connecting the span gas tubing to the inlet fitting or the selection gas valve port Allow the gas to purge through the fitting for about 20 seconds before you tighten it g Press the ALARM BYPASS button and adjust the time displayed ...

Страница 10: ...to a suitable vent Sample gas The sample gas inlet pressure should be between 1 40psig Span gas Span gas if desired must be provided by a user supplied valve Exhaust The exhaust line may be left open or vented to a suitable vent If used with a scavenging system use a large diameter pipe for example pipe as the input to the scavenging system and allow the vent to terminate a little way inside this ...

Страница 11: ...By failsafe is meant that the relays are powered when NOT in alarm so that if power fails they indicate an alarm condition The contacts can handle AC or DC voltages and can carry up to 5A of current for a resistive load Inductive loads such as solenoid valves should be snubbed we suggest that you connect diodes or Zener diodes or Transzorbs directly across them to absorb the inductive spike Do not...

Страница 12: ...ure within this range Higher pressure applications or varying pressure applications will need a regulator to control pressure The analyzer is not sensitive to flow changes between about 0 2SCFH and 5 SCFH but it is sensitive to back pressure changes changes in exhaust pressure The flowmeter is mounted on the exhaust so that the oxygen reading is not affected by potential leaks around the flowmeter...

Страница 13: ... through its O rings The span gas pressure should also be held below 100psig If you permanently connect a span gas tank to it via a user supplied three way valve make very sure that the connection is leak tight as otherwise not only will the calibration be in error but also you will lose all of your span gas as it leaks out Leaks If you suspect a leak due to high oxygen readings change the flow ra...

Страница 14: ...will contaminate the gas in the tank Connect the regulator to the analyzer with a stainless steel line and leak check the connection with Snoop or similar leak detection fluid Purge the line for several minutes with a small flow of gas prior to doing this and leave the gas flowing while you make the connection to the analyzer This prevents a slug of air from giving you excessively high readings wh...

Страница 15: ...ecause that air calibration is in fact very valid You will have to trouble shoot whatever has happened with your gas Output range concept The analyzer displays the oxygen level in appropriate units on the LCD automatically adjusting its sensitivity as required Meanwhile the analog output and the alarms are set on a single user selectable Output range For example you can set the analog output to co...

Страница 16: ...n If the output range does not change the security level must be set to full or span only security In this case change the security level with the laptop and the AMI User Interface program Alarm Set Points The alarm set points can be viewed and changed from the front panel View Alarm Set Points Press either of the ALARM SET POINT buttons and let go The alarm set point will be displayed for about 3...

Страница 17: ... up slowly over the life of the sensor When the value has gotten up to 1000 the sensor has reached the end of its life and should be replaced Also if the value suddenly jumps it indicates that the sensor is getting close to the end of its life Read the Temperature Press the DOWN arrow button The display will show the temperature of the cell block in degrees Fahrenheit The value is limited to 25F a...

Страница 18: ...bration is usually satisfactory Use in a particularly aggressive environment may degrade the sensor faster in this case calibrate more often Sensor Replacement This should be done based on the Span Factor feature rather than as a response to a dead sensor See the chart below for recommended sensor replacement Sensor Part number Description Expected life T2 4SEN09 1 Trace oxygen CO2 background 9 mo...

Страница 19: ... it does immediately flush the affected area with water for a period of at least 15 minutes Refer to the Material Safety Data Sheet provided Dispose of leaking or used sensors in accordance with local regulations Sensors usually contain lead which is toxic and should generally not be thrown into ordinary trash Refer to the MSDS to learn about potential hazards and corrective actions in case of any...

Страница 20: ...lly open the bag using a pair of scissors or a knife Make sure you don t cut yourself or stab the sensor In the rare event that the sensor has leaked there will be liquid in the bag If so do not proceed you need a new sensor Be careful that you don t poke anything such as a fingernail through the membrane 8 Don t pull out the shorting tab yet 9 Holding the sensor by its tab membrane side down slid...

Страница 21: ... to Bleed the regulator first 1 Install the regulator on the span gas bottle but do not open the bottle valve yet 2 Close the regulator exit valve 3 Close the regulator to what would be no pressure on the outlet 4 Briefly open the bottle valve and close it again 5 Loosen the nut connecting the regulator to the bottle and allow the pressure to bleed off and then tighten it again 6 Open the regulato...

Страница 22: ... soap solution or equivalent 4 Verify that the gas flow rate is correct 0 1 to 2 SCFH 5 Oxygen diffusion can be a serious problem Verify that no plastic tubing or other plastic components are used in a trace gas system including diaphragms of pressure regulators packing of valves etc For percent applications similar problems may be experienced with silicone tubing Use Teflon or Tygon or similar hi...

Страница 23: ... sensor replacement instructions under Maintenance No voltage or current output to recording device 1 Verify that the output wires are properly stripped and connected 2 Verify the connections on the output terminal block 3 Verify that the output connections are not shorted all the way back to the recording device Disconnect the wires from the analyzer and use an ohmmeter to check for shorts or ope...

Страница 24: ...1 psig to 100psig Humidity 95 non condensing Operational conditions Pollution degree 2 Installation category I I Drift 1 of full scale in 4 weeks at constant temperature dependent on sensor Expected cell life 9 months to 2 years Response times 90 of full scale in less than 0 10 ppm 25 sec 0 100 ppm 10 sec 0 1000 ppm 10 sec Output 4 20mA isolated Alarm contacts 230 117VAC 5A or 28VDC 5A resistive P...

Страница 25: ...s or associated components AMI s obligation is expressly limited to the analyzer itself The AMI analyzer is not designed as a primary safety device that is to say it is not to be used as the primary means of assuring personnel safety In particular it is not designed to act as a medical instrument monitoring breathing air for correct oxygen concentration and should not be used as such when it is th...

Отзывы: