Sample Handling:
Figure 4 Flow Schematic of the 2001LC
The analyzer expects to get a sample of gas at a pressure between 1 and 100psig. A
built-in needle valve and flowmeter allow the user to control flow to 1SCFH with any
pressure within this range. Higher pressure applications, or varying pressure
applications, will need a regulator to control pressure. The analyzer is not sensitive to
flow changes between about 0.2SCFH and 5 SCFH, but it is sensitive to back pressure
changes (changes in exhaust pressure).
The flowmeter is mounted on the exhaust so that the oxygen reading is not affected by
potential leaks around the flowmeter tube.