14 | Model ta3000R Gas Purity Monitor
Carrier Gas Supplies (On-Line Installation)
Inlet Pressure Range
70 to 90 psig (4.8 to 6.1 bar)
Inlet Pressure Stability
± 2%, regulator required
Return Pressure
Atmospheric vent is optimal, ± 0.5 psig
maximum
Flow Rate
50 cc/min minimum, bypass at 50
cc/min.
Temperature
60 to 100 °F (16 to 38 °C), optimum when
maintained
±
2°C
Maximum Impurity Levels Varries by application. An external purifier
may
be
required
Gas Ports
Sample Inlet
1/16-inch VICI compression fitting
Carrier
1/16-inch VICI compression fitting
Sample Vent
1/16-inch VICI compression fitting
Aux
1/16-inch VICI compression fitting
Sample Gas
Inlet
Fitting
1/16-inch
VICI
compression
fitting
Flow Rate
20 - 100 sccm minimum
Inlet pressure stability
± 2%, UHP regulator required
Vent pressure
Atmospheric pressure vent is optimal, ± 0.5
psig
maximum
Calibration Gas
Inlet Fitting
1/16-inch VICI compression fitting
(Uses
Sample
Inlet)
Cylinder concentration
H
2
, CO, Balance nitrogen, depends
on level of detection
Cylinder volume
100 cubic feet minimum (2800L)
Chassis
Dimensions
7” (18cm) H x 16.8” (43cm) W x 26.5”
(67cm)
D
Weight
35
lb.
(15.9
kg)
Power
100 - 120 VAC, 50/60 Hz; 200-240 VAC,
50/60
Hz
Содержание ta3000R
Страница 43: ...User Manual 43 DIMENSIONAL DRAWINGS ta3oooR Top View ...
Страница 44: ...44 Model ta3000R Gas Purity Monitor 22 115 6 92 ta3oooR Side View ...
Страница 45: ...User Manual 45 24 070 16 70 Outer Dimensions 3D View ...
Страница 46: ...46 Model ta3000R Gas Purity Monitor This page intentionally left blank ...