User Manual | 29
Warm-up
The Power-up screen will display two menu items. Pressing
Warmup
will start the instrument in the warm-up cycle. This cycle will put the
instrument in a ready state for analysis. Pressing
Page
will take you to the
First of 3 system setup pages (covered later).
After pressing
Warm-up
the instrument will perform a series of system
tests and will begin heating critical zones. The results from these tests will
appear on the Warm-up screen. The
Warmup
menu item will also change
to Cool. If you wish to stop the Warm-up process, press the
Cool
key and
the instrument will start to cool down. Do not stop the carrier gas until
all the measured temperatures reach below 50
0
C. These temperatures are
displayed on the Temperature and Staus screen. If any errors occur during
the Warm-up process the system will stop the Warm-up process. Proceed
to Page 41 - error messages for more detail.
Cool Mode
12:34:25
Page
Ver. 2.70
Warmup
Warmup Mode
12:35:25
Page
Module Electronics: Pass
Detector System:
Detector Temperatures:
Carrier Pressure:
Cool
WARM-UP
Power On Screen
Warm-Up Screen