Single channel configuration
This mode displays all the properties of the measurement channel. User can switch to this mode to get all the required information of the selected
channel.
NOTE:
1. If any one of the sensor/multiple sensors used in the formula is in OverRange/UnderRange condition, then an “ERROR” status is displayed in
RED color instead of the current reading. The bar will be drawn with full scale in RED color.
2. If any of the sensors is not connected or not notified, then an “ERROR” status will be displayed for all the channels using that particular sensor
in their formula. The bar will be drawn with full scale in RED color.
3.
OSError:
This error occurs when one or more Linear Sensors used in the formula for the corresponding channel reports “OverSpeed Error”.
When this error occurs the bar is drawn with full scale in RED color. To reset this error, select the channel (in case of multi-channel
configuration scroll the cursor to that channel) and then press ZERO key. It will reset the error and note that the reading from Linear Sensor is
automatically re-zeroed. However it will not re-adjust (or re-Zero) the displayed reading if the channel is in ZERO mode. This operation has to
be explicitly performed by the user through ZERO key.
Measurement Mode:
Can take values
TRACK, PEAK+, PEAK-
Operation Mode:
Can take values
ABS, TARE, PRE
Calculated Reading
Units of measure will be the units
defined in the Computed
Measurement Menu (Section 8.5)
Upper and lower limit
values
Configured Expression/Formula
Channel Number varies from:
C1 to C31