10

Uploading Custom Channels

To upload a custom channel, begin by connecting the 

r

2

i300

 to a computer as described under 

Computer Interface and follow the steps below. 

- Establish communications between the Computer and the   

   

r

2

i300. 

Press the “PC Com Connect” key found in the Reichert  

   

r

2

i300 

PC Interface window. 

To complete the link between the 

   instrument and computer advance to the menu option “PC Com  

   Linkup” and press            to activate. If successful the red light  

   in the software interface window will turn green and the remaining operation picks will 

   become active.

- In the software interface window click on “Custom Channel”.

- Click on Upload Channel and select a channel location by clicking on one of the available  

  channel locations and click on OK.  

- Browse through your directories to locate the desired custom channel file to select it and  

  select open to load file. If successful a message will appear stating “Custom Channel Upload  

  Successful”.

Deleting Custom Channels

- Follow the steps above to establish communications between the instrument and the 

r

2

i300

.

- In the software interface window click on “Download Last Reading Log” and create a txt file  

  where the log will be recorded.

- Once the download is complete and successful click on “Custom Channel”.

- Click on “Delete Channel”, select the channel or channels that are to be deleted and click on  

  Delete to complete the process.

PC Com Linkup

 

Brix-TC                            Brix

Menu

Menu Options

Exit Menu – 

Pressing            when this is displayed will exit the menu mode.

Print Reading Log – 

Provided a printer is connected to the instrument, this feature allows the 

stored readings to be printed. When “Print Reading Log” is displayed and            is pressed the 

entire reading log will be printed, 1 to 1000 readings.

Delete Reading Log – 

Will delete stored values when            is pressed.

Calibration – 

(Covered in greater detail under calibration options) Displays options of: 

1. 

Set (Water) - Pressing            will prompt to 

Apply Water

 then to press            or            .

2. 

Span 1.402XX – See above instructions but 

Apply 1.402XX,

 NIST certified calibration oil  

    when prompted.

3. 

Span 1.514XX –  See above instructions but  

Apply 1.514XX,

 NIST certified calibration oil      

    when prompted.

4. 

Add/Delete/Edit – When selecting any of these modes a “Warning” will appear “Incorrect  

    use of this function can corrupt the instrument”. When changing or adding a set value you  

    must update the entire information string to assure that the instrument will calibrate with  

    proper values. Refractive index, temperature and temperature coefficient information must be  

    known and correct to prevent corruption.  

5. 

Exit -  Press           to exit the calibration options mode.

Contrast – 

Press           to enter and either            or            to adjust display contrast.

PC Com Linkup – 

Communications for this feature is at 38.4K BPS. The software must be 

installed on the PC that the refractometer is connected to and the  “PC Com Connect” button 

must also be pressed on the PC Display for this function to operate. Reference Computer 

Interface section for greater detail.

Set Clock – 

Allows for setting of the time and the date in MDY format.

Select Channel –

 This feature is visible only when at least one custom channel has been 

uploaded to the 

r

2

i300

. Reference the Custom Channels section for greater detail.

Reading Mode –

 Allows changing between available reading modes. TC modes are

 

Temperature Compensated based upon Sucrose data in channel 1. Optional modes in 

channel 1 are nD, nD-TC, Brix or Brix-TC.

Read

Calibrate

Calibrate

Menu

Menu

Menu

Menu

Menu

Menu

Menu

Содержание Reichert r2i300

Страница 1: ...i300 REFRACTOMETER USER GUIDE...

Страница 2: ......

Страница 3: ...recommends that daily calibration with distilled water be performed as a minimum requirement To assure the longest life of your r2 i300 always store the instrument with the prism surfaces and instrume...

Страница 4: ...n 6 Set Point Calibration 6 Multi Point Calibration 6 Add Delete Edit Calibration Points 7 Operational Parameters Sample Temperature Effects on Readings 7 Automatic Temperature Correction TC 7 Compute...

Страница 5: ...power input connector or the unit s power cord Damage to the r2i300 instrument and or injury to the operator may occur Warning The r2i300 should be used in strict accordance with the instructions outl...

Страница 6: ...Voltage 100 240VAC 47 63HZ 1 35A Power Output 40W max 5 VDC 8A Fuse Rating 500mA 250V 5 x 20mm quick acting Power Cords US EU UK supplied Dimensions 12 L x 8 75 W x 3 25 H 30 9cm x 22 5cm x 8 4cm Net...

Страница 7: ...Intrument Features Calibration Oil Sample Well Power Input Connector Power Switch Fuse Holder Com1 PC DB9 Female Com2 Printer DB9 Male Back Panel 4 Touch Screen Display Menu Key Read Key Calibrate Key...

Страница 8: ...e on position for a minimum of one hour before attempting calibration or to measure samples Fig 1 5 Sample Analysis Cleanliness sample quality and sample temperature equilibration are essential to obt...

Страница 9: ...Water appears press and follow the displayed instructions to Apply distilled water Allow adequate time for the sample to temperature equilibrate and press Clean and dry the sample surface and well an...

Страница 10: ...iewing instrument readings One standard method is to have programmed custom channels that calculate the correct reading at 20 degrees C for a sample that is being read at other temperatures The use of...

Страница 11: ...of the instrument through the Menu Options and activate that option To complete the link between instrument and computer click on the Connect PC Comm pick found in the open software window If success...

Страница 12: ...he supplied PC interface software The following describes the r2 i300 custom channel possibilities Channel 1 ICUMSA BRIX The r2 i300 is pre programmed at the factory with channel 1 to read Sucrose or...

Страница 13: ...Reading Log is displayed and is pressed the entire reading log will be printed 1 to 1000 readings Delete Reading Log Will delete stored values when is pressed Calibration Covered in greater detail und...

Страница 14: ...suitable solvent to avoid possibly scratching the prism Use caution when removing samples or drying the prism Care and Cleaning Preparing the prism and well surfaces for samples is essential if consis...

Страница 15: ...s in the refractometer may be required Warranty and Non Warranty Service The warranty period for this instrument is two years against manufacturing defects Please visit our web site at www reichertai...

Страница 16: ...8955 Phone 716 686 4500 email reichertai refractometers ametek com www reichertai com AMETEK GmbH Business Unit Reichert Carl von Linde Strasse 42 85716 Unterschleissheim Munich Germany Tel 49 89 315...

Отзывы: