9

Printer Interface

- Using the supplied Null modem cable connect the cable to the rear of the printer and the  

  instrument where indicated by COM2/PRINTER on the back panel.

- Connect the power supply to the printer and connect to an appropriate power source.

- Turn printer on by positioning the switch to the ON position, switch is found on left side of  

  printer.

- If the light on the front of the printer is not green, press the ON LINE button on the printer to  

  activate the printer.

- To print, access the “Print Reading Log” option and press           .

- A formatted reading will be sent through the Com2/Printer port    

  each time            is pressed.  

Print Reading Log

 

Brix-TC                              Brix

Menu

Read

Custom Channels

The 

r

2

i300

 has the capabilities to measure additional products using custom channels. These 

channels represent scales that have been developed to measure a specific product. Optional 

channels can be selected only after they have been uploaded from a PC to the 

r

2

i300 

using the 

supplied PC interface software. The following describes the 

r

2

i300

 custom channel possibilities. 

Channel 1 – ICUMSA BRIX

The 

r

2

i300

 is pre-programmed at the factory with channel 1 to read % Sucrose or the “Brix” 

scale per the International Commission for Uniform Methods of Sugar Analysis (ICUMSA). 

Temperature Corrected (TC) readings in channel 1 are valid for Brix Solutions only! This 

channel cannot be altered.

Channel 2 – 8, User Programmable 

User programmable, or custom channels as they are commonly 

referred to may be programmed using your own refractive index, 

temperature and percent solids data. Custom channels can 

include definitions for units of measure such as specific gravity, % 

protein, g/ml, etc. Provided a custom channel has been uploaded 

to the 

r

2

i300, 

to access and activate a stored custom channel 

advance to the “Select Channel” menu option and press           . Press            or           to 

advance through the available custom channels and press           to activate.

To learn more about custom channel programming, request “Custom Channel Generation for 

Automatic Refractometer Line” from Reichert Analytical technical support.

Select Channel

 

Brix-TC                              Brix

Menu

Menu

Calibrate

Read

Содержание Reichert r2i300

Страница 1: ...i300 REFRACTOMETER USER GUIDE...

Страница 2: ......

Страница 3: ...recommends that daily calibration with distilled water be performed as a minimum requirement To assure the longest life of your r2 i300 always store the instrument with the prism surfaces and instrume...

Страница 4: ...n 6 Set Point Calibration 6 Multi Point Calibration 6 Add Delete Edit Calibration Points 7 Operational Parameters Sample Temperature Effects on Readings 7 Automatic Temperature Correction TC 7 Compute...

Страница 5: ...power input connector or the unit s power cord Damage to the r2i300 instrument and or injury to the operator may occur Warning The r2i300 should be used in strict accordance with the instructions outl...

Страница 6: ...Voltage 100 240VAC 47 63HZ 1 35A Power Output 40W max 5 VDC 8A Fuse Rating 500mA 250V 5 x 20mm quick acting Power Cords US EU UK supplied Dimensions 12 L x 8 75 W x 3 25 H 30 9cm x 22 5cm x 8 4cm Net...

Страница 7: ...Intrument Features Calibration Oil Sample Well Power Input Connector Power Switch Fuse Holder Com1 PC DB9 Female Com2 Printer DB9 Male Back Panel 4 Touch Screen Display Menu Key Read Key Calibrate Key...

Страница 8: ...e on position for a minimum of one hour before attempting calibration or to measure samples Fig 1 5 Sample Analysis Cleanliness sample quality and sample temperature equilibration are essential to obt...

Страница 9: ...Water appears press and follow the displayed instructions to Apply distilled water Allow adequate time for the sample to temperature equilibrate and press Clean and dry the sample surface and well an...

Страница 10: ...iewing instrument readings One standard method is to have programmed custom channels that calculate the correct reading at 20 degrees C for a sample that is being read at other temperatures The use of...

Страница 11: ...of the instrument through the Menu Options and activate that option To complete the link between instrument and computer click on the Connect PC Comm pick found in the open software window If success...

Страница 12: ...he supplied PC interface software The following describes the r2 i300 custom channel possibilities Channel 1 ICUMSA BRIX The r2 i300 is pre programmed at the factory with channel 1 to read Sucrose or...

Страница 13: ...Reading Log is displayed and is pressed the entire reading log will be printed 1 to 1000 readings Delete Reading Log Will delete stored values when is pressed Calibration Covered in greater detail und...

Страница 14: ...suitable solvent to avoid possibly scratching the prism Use caution when removing samples or drying the prism Care and Cleaning Preparing the prism and well surfaces for samples is essential if consis...

Страница 15: ...s in the refractometer may be required Warranty and Non Warranty Service The warranty period for this instrument is two years against manufacturing defects Please visit our web site at www reichertai...

Страница 16: ...8955 Phone 716 686 4500 email reichertai refractometers ametek com www reichertai com AMETEK GmbH Business Unit Reichert Carl von Linde Strasse 42 85716 Unterschleissheim Munich Germany Tel 49 89 315...

Отзывы: