Reports the contents of the true-time counter in units of 20 ms (50 ticks/s). See also CLEAR_COUNTERS and
SET_TRUE.
Responses:
$G0000000000075<CR>
True time reported as 0 ticks.
$G0000000001076<CR>
True time reported as 1 tick (20 ms).
...
...
$G4294967295132<CR>
True time reported as 4294967295 ticks (over 23000
days).
SHOW_TRUE_PRESET
Reports the current true-time preset in units of 20 ms (50 ticks/s). See also CLEAR_PRESETS and
SET_TRUE_PRESET.
Responses:
$G0000000000075<CR>
True time preset reported as disabled.
$G0000000001076<CR>
True time preset reported as 1 tick.
...
...
$G4294967295132<CR>
True time preset reported as 4294967295 ticks.
SHOW_ULD
Returns the value of the ULD in channels, as a $C record.
SHOW_UNCERTAINTY
Returns the current value of the uncertainty for the peak in the uncertainty preset. See also
SET_UNCERTAINTY.
Responses:
UNCE 0000000000008.5
Uncertainty of the peak is 8.5%.
SHOW_UNCERTAINTY_PRESET
Returns the current uncertainty preset setting. See also SET_UNCERTAINTY_PRESET.
Responses:
UNCE_PRES 000000000000000
No preset.
UNCE_PRES 0000000000008.5
Preset set to 8.5%.
SHOW_VERSION
Reports the EASYNIM-928 firmware version number in the form:
Fmmmm-vvv<CR>
where mmmm is a 4-character model designator and vvv is a 3-character version designator.
Example Responses:
$FDSPR–002<CR>
EASY-NIM-928 firmware Version 2 reported.
SHOW_WINDOW
Reports the start channel and number of channels in the window of interest, in the form
$Dxxxxxyyyyyccc<CR>
where
xxxxx
is the start channel (0–16383) and
yyyyy
is the number of channels (1–16384). See
SET_WINDOW for more information about the window of interest.