124945 01
02-10-2007
29
MGC-HIGH - continued
Wetted parts
:
Aluminium, brass, stainless steel
Leakage
Max. 3% over 5 minutes
Reservoir volume :
500 cm
3
Страница 1: ...User Manual Portable Pressure Calibration Set MGC LOW MGC HIGH Copyright 2007 AMETEK Denmark A S...
Страница 2: ...scribed and accessories may be ordered from the list of accessories Safety symbols This manual contains a number of safety symbols designed to draw your attention to instructions which must be followe...
Страница 3: ...tructions 11 4 1 Operating the MGC LOW 11 4 1 1 Media to be used 13 4 1 2 After calibration 14 4 2 Operating the MGC HIGH 15 4 2 1 Media to be used 18 4 2 2 After calibration 18 4 3 Operating the opti...
Страница 4: ...tion This expertise is reflected in our products which are all designed for daily use in an industrial environment Please note that we would be very interested in hearing from you if you have any idea...
Страница 5: ...hnical assistance Please contact the dealer from whom you acquired the pressure calibration system if you require technical assistance Guarantee According to current terms of sale and delivery This gu...
Страница 6: ...n those described in this manual and for any application other than precision pressure calibration jobs The pressure system should only be used by TRAINED PERSONNEL Never subject the pressure gauges t...
Страница 7: ...product liability only applies if the pressure system is subject to a manufacturing defect This liability becomes void if the user fails to follow the maintenance instructions set out in this manual o...
Страница 8: ...0 40 BAR Kg cm 0 100 BAR Kg cm and 0 300 BAR Kg cm System 3 0 10 BAR Kg cm 0 40 BAR Kg cm and 0 100 BAR Kg cm System 4 0 10 BAR Kg cm 0 40 BAR Kg cm and 0 300 BAR Kg cm Both models are supplied with...
Страница 9: ...against the list shown below If any of the parts are missing or damaged please contact the dealer who sold the pressure calibration system You should receive 1 pump system T 620 incl 0 5 m hose MGC L...
Страница 10: ...10 02 10 2007 124945 01 3 adapters 1 4 x 1 4 BSP 1 2 x 1 4 BSP 1 4 x 3 4 BSP MGC HIGH 1 plug 1 4 BSP 2 bonded seal 1 4 BSP 1 traceable certificate per gauge 1 user manual...
Страница 11: ...INED PERSONNEL Never subject the pressure gauges to overpressure In case of overpressure the gauge should not be used for calibration purposes None of our calibration systems are cleaned or prepared f...
Страница 12: ...pressure should be indicated If this is not the case the pressure gauge may have been subjected to over pressure and should not be used for calibration purposes 2 SYSTEM 1 0 6 Kg cm2 0 16 Kg cm2 0 16...
Страница 13: ...be released from the system by loosening the connectors pos 5 It is now possible to pump out oil and air until no air is left in the oil 4 If other connectors than those supplied are needed please use...
Страница 14: ...KYDROL or the like 4 1 2 After calibration 1 Open the pump valve pos 1 2 Make sure that the pressure system is depressurised before all items are disconnected 3 Disconnect all items 4 Close the pump v...
Страница 15: ...L Never subject the pressure gauges to overpressure In case of overpressure the gauge should not be used for calibration purposes None of our calibration systems are cleaned or prepared for OXYGEN OBJ...
Страница 16: ...is is not the case the pressure gauge may have been subjected to over pressure and should not be used for calibration purposes SYSTEM 1 0 6 Kg cm2 0 16 Kg cm2 0 10 Kg cm2 0 40 Kg cm2 0 10 Kg cm2 0 40...
Страница 17: ...rs pos 6 It is now possible to pump out oil and air until no air is left in the oil 4 If other connectors than those supplied are needed please use fittings with correct pressure rating and correct th...
Страница 18: ...1 Open the pump valve pos 2 on the right side of the pump 2 Open the pump valve pos 3 on the front of the pump gently to protect the gauges from the pressure shock 3 Make sure that the pressure system...
Страница 19: ...any purposes other than those described in this manual Do not use this system for any applications other than precision pressure calibration jobs Never subject the valve system to overpressure Do not...
Страница 20: ...stems with gauges are prepared for test fitted with a valve with test connector To operate this kind of system a Lloyds valve system with clamp from AMETEK Denmark A S can be used Pressure system with...
Страница 21: ...lve system pos 3 5 Connect the Lloyds valve system to the valve with test connector located on the system which is to be tested pos 4 6 Open the valve on the system to be tested pos 5 7 Open the Lloyd...
Страница 22: ...TEK Denmark s liability ceases if parts are replaced repaired using spare parts which are not identical to those recommended by the manufacturer non original parts are used in any way when operating t...
Страница 23: ...before the pressure is stable enough for calibration Release air See Notes page 13 or 17 The effect will disappear rapidly The effect will disappear rapidly Note Always readjust the pressure after 5 1...
Страница 24: ...set to the manufacturer for service please enclose a fully completed service information form Simply copy the form on the following page and fill in the required information The pressure calibration s...
Страница 25: ...________________ Pressure Service request This instrument is sent for calibration please tick off ___ Calibration as left ___ Check ___ Calibration as found and as left ___ Service ___ Accredited cali...
Страница 26: ...wing cleaning procedures as and when required The exterior of the instrument Clean using water and a soft cloth The cloth should be wrung out hard to avoid any water penetrating the calibrator and cau...
Страница 27: ...W Range 0 6 BAR kg cm 0 10 BAR kg cm 0 16 BAR kg cm 0 40 BAR kg cm 0 100 BAR kg cm 0 300 BAR kg cm Max pressure 350 BAR 5000 PSI Min pressure 0 BAR MGC HIGH Range 0 6 BAR kg cm 0 10 BAR kg cm 0 16 BAR...
Страница 28: ...p 0 to 50 C 32 to 122 F Humidity range 0 to 90 RH Protection class IP33 Wetted parts Aluminium brass stainless steel polycarbonat Leakage Max 5 over 5 minutes Reservoir volume 130 cm3 full 180 cm3 MGC...
Страница 29: ...124945 01 02 10 2007 29 MGC HIGH continued Wetted parts Aluminium brass stainless steel Leakage Max 3 over 5 minutes Reservoir volume 500 cm3...
Страница 30: ...2 Gauge System 4 0 10 BAR kg cm 2 MGC LOW 126319 Gauge System 4 0 300 BAR kg cm 2 MGC LOW 126320 Pump system P 142 System 1 2 3 and 4 MGC HIGH 103306T4 Gauge System 1 0 6 BAR kg cm 2 MGC HIGH 100403 G...
Страница 31: ...60R120 1 0 m 3 3 ft hose with 1 4 BSP female male MGC LOW 65P175 2 0 m 6 6 ft hose with 1 4 BSP female male MGC LOW 65P180 1 5 m 4 92 ft hose with 1 4 BSP female male 60I156 5 0 m 16 4 ft hose with 1...