Page 22
LSP Ex System for Zone 22
Linescanner Head
User Guide
Cable limitations on LSP Head performance
The LSP Head is a fast response precision instrument. However, some cable
configurations may limit the practical (useable) scanning speed and the dynamic
performance of the LSP Head, through cable capacitance and external noise.
The following tables show the recommended cables when considering cable
length and scan frequency.
Total cable
distance
(scanner
head to
processor
LSP
Head
Cable 1
Intermediate
connection
Cable 2
Processor
Maximum
recommended scan
speed (Hz)
LSP Ex
1/2
LSP Ex
5/6/7
Up to 15m
Any
* LSP cable 15m
LSC-C
100
100
Up to 65m
LSP cable 15m
Terminal box with
PSU (802662)
LSP type cable
50
100
LSP cable 15m
Terminal box with
PSU (802662)
Belden 8162 /
8163 & 8777
100
100
Up to 165m
LSP cable 15m
Terminal box with
PSU (802662)
Belden 8162 /
8163 & 8777
50
100
Up to 315m
LSP cable 15m
Terminal box with
PSU (802662)
Belden 8162 /
8163 & 8777
20
50
Up to 15m
Any
* LSP cable 15m
LSC-B
100
100
Up to 50m
* LSP cable 15m
20
100
Up to 65m
LSP cable 15m
Terminal box
(802869)
LSP type cable
50
100
LSP cable 15m
Terminal box
(802869)
Belden 8162 /
8163 & 8777
100
100
Up to 165m
LSP cable 15m
Terminal box
(802869)
Belden 8162 /
8163 & 8777
50
100
Up to 315m
LSP cable 15m
Terminal box
(802869)
Belden 8162 /
8163 & 8777
20
50
Notes
1
LSP 15m cable can be standard 031-402 or high temperature 031-794
2
LSP type cable is 12 core 16/0.2mm DEF STAN 61.12pts 4 & 5
3
Terminal box 802869 is a passive connector allowing connection of different
cable types
4
Terminal box with PSU 802662 includes a local LSP head power supply. This
is used where the head to processor cable distance exceeds 15m when using
an LSC-C processor, and 165m when using an LSC-B processor
5
Belden cable 8162/8163 is used for the Scan Valid and Fast temperature
signals from the LSP head
6
Belden cable 8177 may be used for additional signals such as Emissivity,
IATS and remote laser switch