Page 19
User Guide
LSP Ex System for Zone 22
Linescanner Head
Hotspot IR 5, 6 & 7 Ex Target size table
The nominal Target Size (TS) is given by:
•
At target distances less than 1200mm / 47 in:
TS = 12mm / 0.47in
or
•
At target distances greater than 1200mm / 47 in:
TS = TD
FoV
•
Hotspot IR 5 Ex, Hotspot IR 6 Ex & Hotspot IR 7 Ex FoV=100:1
TD
SW
TP
Hotspot IR 5, 6 & 7 Ex - Target Size
Target Distance
Scan Width Target Path 0° Scan angle
+/-40° (x axis)
+/-40° (y axis)
mm
in
mm
in
mm
in
mm
in
mm
in
mm
in
200
7.9
394
15.5 261
10.3 12.0 0.47
15.7 0.62
12.0 0.47
500
19.7
898
35.4 653
25.7 12.0 0.47
15.7 0.62
12.0 0.47
750
29.5
1317 51.9 979
38.5 12.0 0.47
15.7 0.62
12.0 0.47
1000 39.4
1737 68.4 1305 51.4 12.0 0.47
17.0 0.67
13.1 0.52
1200 47.2
2073 81.6 1566 61.7 12.0 0.47
20.4 0.80
15.7 0.62
1500 59.1
2576 101.4 1958 77.1 15.0 0.59
25.5 1.00
19.5 0.77
2000 78.7
3415 134.5 2611 102.8 20.0 0.79
34.0 1.34
26.0 1.02
3000 118.1
5093 200.5 3916 154.2 30.0 1.18
51.0 2.00
39.0 1.54
4000 157.5
6772 266.6 5222 205.6 40.0 1.57
68.0 2.68
52.0 2.05