Table of contents:

Page

1.0

General information ......................4

2.0

Introduction....................................5

3.0

Functionality ..................................6

3.1

Functional description..................6

3.2

Connections ................................6

3.3

Serial number..............................8

4.0

Operation........................................9

4.1

Operation area ............................9

5.0

Maintenance.................................10

5.1

Calibration .................................11

6.0

Technical specifications .............15

Содержание JOFRA RTC-156/157 B/C

Страница 1: ...User manual JOFRA DLC 155 158 250 700 Copyright 2010 AMETEK Denmark A S...

Страница 2: ......

Страница 3: ...age 1 0 General information 4 2 0 Introduction 5 3 0 Functionality 6 3 1 Functional description 6 3 2 Connections 6 3 3 Serial number 8 4 0 Operation 9 4 1 Operation area 9 5 0 Maintenance 10 5 1 Cali...

Страница 4: ...ion This manual is only effective for the following products JOFRA DLC 155 JOFRA DLC 158 JOFRA DLC 250 JOFRA DLC 700 The products are manufactured by AMETEK Denmark A S Gydevang 32 34 3450 Aller d Den...

Страница 5: ...temperature measuring with JOFRA RTC 156 157 B C JOFRA RTC 158 B C JOFRA RTC 250 B C and JOFRA RTC 700 B C systems and are ready for use Please read this manual carefully before use to obtain maximum...

Страница 6: ...C to 155 C 76 F to 311 F DLC 250 0 C to 250 C 32 F to 482 F DLC 700 0 C to 700 C 32 F to 1292 F The JOFRA DLC 155 DLC 158 DLC 250 and DLC 700 probes may be supplied with a certificate for a limited t...

Страница 7: ...127414 01 20 12 2010 7 The figure below is shown from the connector side of the probe connector 1 2 3 4...

Страница 8: ...8 20 12 2010 127414 01 3 3 Serial number The serial number is placed on the probe as shown on the figure below Serial number...

Страница 9: ...peration 4 1 Operation area All the probes are intended for use in areas which meet the following Ambient temperature range 20 C to 70 C 4 F to 158 F Humidity 0 to 90 Protection class IP 50 Warning Do...

Страница 10: ...user may carry out the following procedure himself Cleaning the probe Use alcohol or water and a soft cloth Caution The probe must always be protected against any mechanical damage The probe must neve...

Страница 11: ...rature probe recommended or similar is required Procedure 1 Place the DLC probe in the insert and connect the probe to the DLC input of the calibrator 2 Go to the Sensor setup menu 3 Set Use load comp...

Страница 12: ...122 F 50 C 122 F 500 C 932 F 100 C 212 F 100 C 212 F 660 C 1220 F Recommended calibration temperatures 155 C 311 F 155 C 311 F 700 C 1292 F Lower measuring point Bottom of insert Bottom of insert Bot...

Страница 13: ...bottom 60 mm 2 4 from bottom Curve fit warning limit 0 01 C 0 018 F 0 01 C 0 018 F Note The temperatures must be entered in Celcius The following procedure is repeated for each temperature 1 Set the S...

Страница 14: ...til the temperature have stabilized and the DLC reading is 0 00 C 6 Record the temperature of the reference probe and the DLC probe in the excel spreadsheet 7 Go to next SET temperature After the last...

Страница 15: ...311 F Diameter OD3 mm Depth 140 mm 5 51 in Media compatibility AISI316 Connection Redel plug with build in memory is standard Probe specifications DLC 158 Sensor type Differential TC type T Probe len...

Страница 16: ...D3 mm Depth 140 mm 5 51 in Media compatibility AISI316 Connection Redel plug with build in memory is standard Probe specifications DLC 700 Sensor type Differential TC type N Probe length 177 mm 6 96 i...

Страница 17: ...ecommended below according to the ITS 90 temperature scale The DLC 155 158 probe is as standard calibrated in the range 45 C to 155 C 49 F to 311 F The DLC 250 probe is as standard calibrated in the r...