The calibration temperatures are:
Sensor
DLC-155
DLC-155
DLC-700
Calibrator
RTC-156
RTC-157
RTC-700
-30°C/-22°F
-45°C/-49°F
33°C/91.4°F
-20°C/-4°F
-20°C/-4°F
200°C/392°F
0°C/32°F
0°C/32°F
350°C/662°F
50°C/122°F 50°C/122°F 500°C/932°F
100°C/212°F 100°C/212°F 660°C/1220°F
*
Recommended
calibration
temperatures
155°C/311°F 155°C/311°F 700°C/1292°F
Lower
measuring point
Bottom of
insert
Bottom of
insert
Bottom of
insert
Upper
measuring point
60 mm /
2.4“ from
bottom
60 mm /
2.4” from
bottom
80 mm / 3.1”
from bottom
Curve fit
warning limit
0.01°C/
0.018°F
0.01°C /
0.018°F
0.05°C /
0.09°F
*
Note :
The temperatures must be entered in Celcius.