2.
The NEXT option can be used to select which display to act upon. The
ON/OFF option turns the selected display on or off. The selected display
and current on/off state are displayed in the lower display.
3.
Use the DONE option to save the changes and return home. When a
display is deactivated its configuration is retained. When the display is
activated its configuration is checked against the configurations of the
other currently active displays, if the configurations are in conflict the
recalled display's configuration is modified to avoid the conflict. If all
three displays are deactivated the LOWER display will come on automat-
ically.
2.4.09 Setting the RTD probe type
The JOFRA HPC600 has a built in high accuracy RTD thermometer, it works
with an RTD sensor (optional).
To select type of temperature sensor proceed as follows:
1.
With the calibrator turned on and operating press the F3 key to activate
the MORE menu option. Press the NEXT button until PROBE TYPE
appears in the left text field. Now press the F1 key to enter the function.
2.
There are four probe types to select from P100-385, P100-392, P100-JIS
and CUSTOM. Use the SELECT option to select the desired probe type
and the DONE option to store the change and return home.
Note
: The default probe type is PT100-385.
3. Pressing DONE leaves the temperature selection function
17