location. Then use the save option to store the current set-up into the
selected location or the recall option to recall the set-up stored in the
selected location. The display menu will automatically go home.
2.4.07 Setting AutoShut-off Parameters
The calibrator can be set to automatically shut-off after a selected number of
minutes; this function can also be disabled.
To change parameters proceed as follows:
1.
With the calibrator turned on and operating press the F3 key to activate
the MORE menu option. Press the NEXT button until AUTO OFF
appears in the left text field. Now press the F1 key to enter the setup.
2.
To set the auto shut off parameters select the AUTO OFF option on the
Auto Shut Off Main Menu.
3.
Use the cursor keys to select the number of minutes before the calibrator
turns off or disable auto shut-off by scrolling all the way down to 0.
4.
Use the AUTO OFF DONE option to set the parameters and return
home. The auto shut off time is reset whenever a key is pressed.
2.4.08 Activating and Deactivating a Display
This is where the number of active measuring windows / channels are
selected, 1, 2 or 3 windows can be selected, to give the optimal mix
between text size, overview and amount of information.
To use the number of windows, proceed as follows:
1. With the calibrator turned on and operating press the F3 key to activate
the MORE menu option. Press the NEXT button until DISPLAY appears
in the left text field. Now press the F1 key to enter the function.