129375 06
2019-12-06
48
Adjusting temperature, internal reference
Adjusting temperature, external reference
Adjusting temperature, internal & external reference
1. Place the reference sensor in the calibrator.
2. Select the Adjust temp., mode for the wanted adjustment procedure.
3. Press (Enter) and the calibrator starts heating up/cooling down to reach the
first calibration temperature.
4. When the calibrator is stable an audible alarm will be emitted.
5. Type in the reference temperature found using the reference thermometer. Use
the (Up) and (Down) keys to type in the value.
6. Press (Enter) to accept the value.
7. Press Continue (F3) to activate the heating up/cooling down process and
wait for the stability indication.
8. Once the temperature is stable, press Continue (F3) again.
9. Continue pressing the (Continue) key until the adjusting process is
completed.
10. When the temperature has been adjusted a new calibration date must be typed in.
The calibrator automatically asks for a new date.
Adjusting temperature, self calibration
Note…
Performing this adjustment will overwrite the above-mentioned adjustment.
1. Place the external JOFRA STS reference sensor in the calibrator and connect it to
the external reference input.
2. Select the Adjust temp., self calibration mode.
3. Press Start (F4). The self calibration starts running automatically step by
step as shown in the table below.