Note…
that when “SET follow true” is selected, the calibrator will control the temperature
to the TRUE temperature. This means that it could take longer time before the
calibrator indicates stability.
5.4
Stability setting
True: Internal reference
When internal reference is selected the calibrator uses a set of minimum internal
stability criteria that shall be met before stability is indicated.
The stability time can be adjusted from the minimum internal stability time (typical 5
min) up to 99 min.
True: External reference
When internal reference is selected the user can adjust both stability tolerance and
stability time.
The Stability tolerance can be set between 0.01°C to 1.00°C. The tolerance should
be set low enough to utilize the good temperature stability of the calibrator –
however a low value also gives a longer time to be stable.
The stability time can be set from 1 – 99 minutes.
When the TRUE temperature has reached the specified Stability tolerance during
the specified Stability time, then the stability indicator in the main screen will turn
green.
5.5
Selecting the set-temperature
Note…
Temperature range is limited by Min. SET Temp. and Max. SET Temp. settings
editable in the System Settings menu and by the temperature range of the
external reference sensor, if connected.
The set-temperature can be entered both manually and by selecting a preset
temperature.
2. For manually use press (Set)
3. Use the (Arrow) keys to enter the temperature requested.
4. Press (Enter) to accept the entered temperature.