2. Press one of the (Arrow) keys to access the editable fields for new values:
•
No of steps: the number of temperature steps per direction (T
1
T
x
) can be
set using integers from 2 to 12. You must select minimum 2
steps, maximum 12 steps.
•
Hold time:
defines the time (in minutes) the temperature is maintained
(after it is stable) for each step.
•
Step values: must be set within the sensors permitted range.
3. Press (F4) to start the Auto Step test.
The Auto Step test is now in progress.
While the Auto Step test is in progress, 4 options are available:
•
Stop :
Press (F1) to stop the Auto Step test. The process will
end.
Pressing the (F4) key the process will start again running
step 1.
•
Previous : Press (F2) to force the test to jump a step backwards to the
previous running step regardless of the step’s stability.
•
Next :
Press (F3) to force the test to jump a step forwards to the
next running step regardless of the step’s stability.
•
Pause :
Press (F4) to pause the test. Pressing start (F4)
again, the process will continue running from the current step.
5.6.2
The calibrator's Auto Step procedure
1) Once the Auto Step test is started, the calibrator starts working towards the given
set-temperature. An audible alarm will be emitted once the calibrator is stable.
2) The calibrator will wait the specified amount of hold time. The instrument indicates
this by counting down the amount of time remaining:
3) The calibrator will then go to the next step. The procedure is the same as for the first
step. This process will be repeated until the last step has been executed and the
function has been completed.