102
23-03-2009
105446 06
THERMAL SPECIFICATIONS
ATC-156 A/B
Influence from load with
Ext. Reference :
0.01°C
(0.02°F)
@155°C (311°F)
0.01°C (0.02°F) @-24°C (-11.2°F)
Long term drift (1 year) :
±0.05°C
(±0.09°F)
Calibration accuracy (test limit)
±0.05°C (0.09°F)
Temperature coefficient
5ppm/°C (0-40°C) / (32-104°F)
Stability
±0.01°C ( ±0.02F)
Reference accuracy
±0.02°C (±0.04°F)
Total accuracy
±0.10°C (±0.18°F)
Heating time incl. insert
-24°C / -11.2°F to 23°C / 73.4°F :
4 min.
23°C / 73.4°F to 100°C / 212°F:
9 min.
23°C / 73.4°F to 155°C / 311°F :
19 min.
100°C / 212°F to 155°C / 311°F:
10 min.
-24°C / -11.2°F to 155°C / 311°F :
23 min.
Time to stability
7 min.
Cooling time incl. insert
155°C / 311°F to 100°C / 212°F:
4 min.
155°C / 311°F to 23°C / 73.4°F :
13 min.
100°C / 212°F to 23°C / 73.4°F:
9 min.
23°C / 73.4°F to 0°C / 32°F:
6 min.
23°C / 73.4°F to –24°C / -11.2°F :
29 min.
0°C / 32°F to –20°C / -4°F:
13 min.
155°C / 311°F to –24°C / -11.2°F : 42 min.