16040-101 Rev. J
25
Instrument Operation
(continued)
Measurement Process
(continued)
10. Move the Control button to the right (
) or left (
) as
needed to acquire measurements on the other eye and
repeat the above process.
11. To review the measurement data, you can “toggle” between
the right and left eye by moving the control button right (
)
or left (
).
12. To send the measurement data to a Bluetooth printer or
computer EMR system, move the Control button down
(
)
.
The data will be sent to the connected device, if the data is not
transferred refer to the Bluetooth setup section of this user’s
guide.
Note:
If a Bluetooth device was found and the connection estab-
lished in Setup, then move the Control button down (
)
to send the data to the Bluetooth device.
13. To clear the measurement data, move the control button
up (
) and hold it until the display shows “measurement
cleared.”
14. Perform the cleaning instructions in the Cleaning and
Disinfection section of this manual.
Note:
If you are unable to take 25 measurements, the average
will be displayed for the number of measurements that
were taken. Either press the control button down once
or wait for about 15 seconds and the display will change
to show the print and clear icon.