AMETEK CTS 

 

Quick start guide - Conducted 

 

1.00 

7 / 50 

 

CAUTION

 

Before using this equipment, read the user manual and the  

separate delivered safety manual carefully 

 

1.3. 

Responsibility of the operator 

These operating instructions form an essential part of the equipment and must be available to the operator at all times. The 

user must obey all safety instructions and warnings. 

 

CAUTION

 

The purpose of this instrument is the generation of defined interferences signals for EMI immunity 

testing. Depending on the arrangement of the test rig, the configuration, the cabling and the properties 

of the EUT itself, a significant amount of electromagnetic radiation may result that could also affect other 

equipment and systems.  
The equipment is designed to operate in industrial environment. For operating in other or sensitive 

environment, such as light industry, airport area…, the user may use a shielded room for operate.  
The user himself or herself is ultimately responsible for the correct and controlled operation of the rig. 

In case of doubt, the tests should be carried out in a Faraday cage. 

 

1.4. 

General hazard 

Before applying power to the system, verify that your product is configured properly for your particular application. 

 

WARNING

 

The generators and their accessories operate at high voltages. 
Hazardous voltages may be present when covers are removed. Qualified personnel must use extreme 

caution when servicing this equipment. 
Circuit boards, test points, and output voltages also may be floating above (below) chassis ground. 

 

The design of external insulation must be such that it exceeds the maximum impulse voltages of the 

generator (5.5 kV or 7.0 kV). 

 

 

Only 

qualified personnel 

who deal with attendant hazards in impulse generators, are allowed to perform installation and 

servicing. 
Ensure that the AC power line ground is connected properly to the Power Rack input connector or chassis. Similarly, other 

power ground lines including those to application and maintenance equipment 

must 

be grounded properly for both personnel 

and equipment safety. 
Always ensure that facility AC input power is de-energized prior to connecting or disconnecting any cable. 
The user must ensure that the output power lines are labeled properly as to the safety hazards and that any inadvertent 

contact with hazardous voltages is eliminated. 

Guard against risks of electrical shock during open cover checks by not touching any portion of the electrical circuits. Even 

when power is off, capacitors may retain an electrical charge. Use safety glasses during open cover checks to avoid personal 

injury by any sudden component failure. 

 

Neither AMETEK CTS GmbH, nor any of the subsidiary sales organizations can accept any responsibility for personnel, 

material or inconsequential injury, loss or damage that results from improper use of the equipment and accessories. 

 

 

Содержание EM TEST compact NX

Страница 1: ...name UserManual Short Conducted E V1 00_EN docx Print date 4 29 2021 QUICK START GUIDE AND SAFETY INSTRUCTIONS Transient Immunity Generators compact NX NSG 3000A And Accessories coupling NX CDN 30x0A...

Страница 2: ...cted 1 00 2 50 AMETEK CTS GmbH Sternenhofstrasse 15 4153 Reinach BL1 Switzerland Phone 41 61 204 41 11 Fax 41 61 204 41 00 URL http www ametek cts com Copyright 2021 AMETEK CTS GmbH All right reserved...

Страница 3: ...s to be taken 15 4 2 Failures and Heavy Strain 15 4 3 Coupling Networks 15 4 4 Interference to the environment 16 5 Testing 17 5 1 Safety Precautions 17 5 2 Danger from EUT 17 6 Quick start guide comp...

Страница 4: ...39 9 5 1 Coupling with 40 resistor and 0 5 F capacitor 39 9 5 2 Coupling with 40 Ohm and Gas Discharge Tube GDT 39 9 6 Coupling to GND 40 9 6 1 Test setup Coupling to PE 40 9 7 Coupling network DCD s...

Страница 5: ...AMETEK CTS Quick start guide Conducted 1 00 5 50...

Страница 6: ...The WARNING symbol indicates a potential hazard It calls attention to a procedure practice or condition which if not followed could possibly cause bodily injured or death If a WARNING is indicated do...

Страница 7: ...resent when covers are removed Qualified personnel must use extreme caution when servicing this equipment Circuit boards test points and output voltages also may be floating above below chassis ground...

Страница 8: ...bination with items articles or materials not authorized by AMETEK CTS The Buyer may not assert any claim that the products are not in conformity with any warranty until the buyer has made all payment...

Страница 9: ...a detailed list of the contents Before activating the equipment check for damage which may have occurred during shipment Check each container as well as the generator itself In case of physical damage...

Страница 10: ...an auto transformer which is connected to a higher supply voltage the base of the auto transformer must be connected to the neutral of the power supply Equipment of Class I which is supplied with a 3...

Страница 11: ...EUT power supply It is in the scope of responsibility of the user to protect the EUT external for the rated current CAUTION The design of the external fuse must be matching the following rules fuse d...

Страница 12: ...e system and enables the generation of the high voltage impulses in the generators Design Each device that has internal relevant high voltage unit includes a safety circuit The safety circuit works li...

Страница 13: ...ase order the optional Safety Circuit Adapter SCC AD Figure 13 SCC AD option for external safety circuit loop Figure 14 Safety Termination 3 2 Warning lamp 3 2 1 Warning lamp for compact NX and NSG 30...

Страница 14: ...t setup installation Figure 18 Earth with Brass Bar 3 4 FI Fault current protection The standards recommend decoupling and filter capacitors to PE for decoupling surge pulses This is the reason for tr...

Страница 15: ...onnected to the ground reference plane of the test setup Please ensure the correct current back flow to the generator Missing connections can cause extremely high voltages and may become a health haza...

Страница 16: ...d RF etc Therefore a disturbance of the environment cannot be excluded The user has an obligation to use a suitable test environment in order to minimize influences to the environment This may require...

Страница 17: ...topped and the EUT disconnected from the voltage power supply The EUT must be tested within a safety container or within a protected area In extreme circumstances the EUT may ignite or explode as a re...

Страница 18: ...2 Power Mains cable Power Mains cable Connectors country coded 3 ESC EUT Supply Cable Connectors country coded 4 ESA 1 EUT Supply Adapter Connectors country coded 5 SCT 112801 Safety Circuit Terminato...

Страница 19: ...t 1 Transducer plate Support PVF AD3 Pulse Verification Fast Adapter 3 MC to SHV ESS 1 111607 ESS 1 Interlock for compact NX5 system Switches OFF High voltage and EUT power supply SCC AD 111240 Safety...

Страница 20: ...ambient temperature or humidity levels Use only officially approved connectors and accessory items Ensure that a reliable return path for the interference current is provided between the EUT and the...

Страница 21: ...ground Figure 20 Grounding and power connection System GND The system is connected to GND via the earth wire of the connected mains power cable for the generator and the EUT supply A separate GND wir...

Страница 22: ...ound connection 7 Switch the system on and operate as instructed in this manual 6 4 5 Connecting the generator to the ground reference The earth connection to the GRP should be low inductive and low o...

Страница 23: ...aker box with 16 A protection is installed between the building supply and the test generator Test generator and EUT are now fused for 16 A rated current Figure 24 External Circuit Breaker 6 6 EUT pow...

Страница 24: ...6 7 Front view model specific distinctions Figure 26 Operating Elements 1 Active indication 2 Touch screen 3 Knob Inc Dec Enter 4 Test On 5 CRO I surge 6 CRO V surge 7 HV pulse Burst output 50 8 Coupl...

Страница 25: ...he generator operation It is in the responsibility of the user to decide to run the test in correct phase neutral connection Anyway with wrong phase connection it is possible and normal to measure a v...

Страница 26: ...AMETEK CTS Quick start guide Conducted 1 00 26 50 Figure 29 Loading Home screen after booting The home screen shows all detected symbols of phenomenon with their color Figure 30 The Home Screen...

Страница 27: ...ections 1 Must be operated only by authorized and trained specialists 2 Do not operate without any ground connection 3 Hazardous high voltages are generated on the output terminal 1 EUT Equipment Unde...

Страница 28: ...en 4 Test and Control Buttons 5 Output and coupling options 1 Each phenomenon is color coded Tap to select 1 Tapping QUICK START will enable you to start a test 2 Pressing either BACK or on the screen...

Страница 29: ...Graphic Press loupe symbol to enlarge or swipe left and right to naviagte to other views 3 Press to change adjustment range Selected range will be shown briefly next to the selected parameter Push and...

Страница 30: ...the EUT 2 Press START PAUSE to start a test 3 Long press START PAUSE for continuous test 1 Press START PAUSE to pause or press STOP to stop a test 2 Press START PAUSE to continue or start a test 3 Aft...

Страница 31: ...that a reliable return path for the interference current is provided between the EUT and the generator The GND wire from the generator to the reference GND and must be designed as a low inductance con...

Страница 32: ...the coupling matrix can be installed one above the other The coupling matrix should be used as central output for the EUT and should be mounted directly onto the ground reference plane The compact NX...

Страница 33: ...4 Connecting CDN 30x3A C32 with NSG 30x0A generator When setting up the test national and international regulations regarding human safety have to be guaranteed Figure 37 Example Test Rack All units t...

Страница 34: ...ion cable BCC 1000 3 Output to the EUT Burst and Surge pulses are coupled to the supply lines 4 Surge output to external CDN for data lines via a HVS cable Figure 3 2 Front side NSG 3060A with CDN 306...

Страница 35: ...5 TEST ON button 6 EFT input from compact NX generator 7 EFT output to capacitive coupling clamp 8 EUT Output L1 DC L2 L3 N DC PE 9 GND reference for EFT verification 10 Output HV COM to external Sur...

Страница 36: ...ler is connected and shows the possible coupling modes in the display of the generator The complete system is controlled by the generator from front panel or software If the external coupling network...

Страница 37: ...n the application the sequence of the earth connection DCD generator reference earth can be changed as required WARNING The device earth bolt must be connected to the ground reference plane if there i...

Страница 38: ...n the various models is Number of data lines DCD sr 4 x series with 4 lines DCD sr 8 x series with 8 lines nominal voltage and current level of the signal and data lines various test voltage level of...

Страница 39: ...ge Tube GDT CAUTION Use only bridges with built in GDT gas discharge tube Otherwise the generator circuit is loaded with the line voltage in series with 40 This can destroy the generator Figure 47 Exa...

Страница 40: ...rk DCD st 4 lines 4 Data line setup Each of the 4 lines has an impedance of 160 All four lines are connected by using a coupling device CD Figure 49 4 Data Line Setup 2 Data line setup Two lines are s...

Страница 41: ...re 53 80 Lines 9 9 Test setup for 10 700 s impulses Due to the nature of the wiring used for unshielded outdoor symmetrical communication lines twisted pairs the coupling is always in common mode The...

Страница 42: ...esults by shorting the 135 resistors Bridge CD Connection for the built in coupling device GDT ABD for coupling to lines L1 to L4 Figure 55 Two Symmetrical Lines 2 Data line setup One symmetrical line...

Страница 43: ...with the signal flow It is often impossible to contact the required circuit direct e g coaxial or shielded cables In this case the coupling is realized by the capacitive coupling clamp The interferenc...

Страница 44: ...tly applied to the metallic case of the DUT The HSC 4 8 may also be used to couple Ring Wave impulses 0 5 us 100 kHz and burst pulses to shielded cables 11 2 CNI 508 N2 Coupling decoupling network The...

Страница 45: ...rom Surge Generator to shield 7 Surge input from Surge Generator to lines 8 Handle closed position 9 Output to RJ45 AE Port 10 Earth plugs 11 5 Front and rear view CNI 508N2 1 Burst input from Burst G...

Страница 46: ...couplings with separate input connectors Surge and Ring wave pulses to data lines Surge pulses to shield Burst pulses to shield For an additional overvoltage protection of the AE port the SPN 508N2 ca...

Страница 47: ...Test setup The SPN 508N1 protects the auxiliary equipment with a residual voltage of less than 10V Generator output plugs The Surge and Ringwave output plugs are located at the rear side of the impuls...

Страница 48: ...l protect the AE port side and all connected devices The surge coupling voltage and current to the lines depends on the impedance connected at the HSC 4 8 AE port In case of a short circuit as shown i...

Страница 49: ...n In case of a high protection level of approx 10V at the AE port the user can connect additional the SPN 508N1 This protection will separate the AE side the surge by a galvanic insulation from the su...

Страница 50: ...must only be performed by qualified service technicians who are trained and familiar with the dangers of servicing the EM TEST TESEQ generator Only fuses of correct voltage and amperage as specified b...

Отзывы: