Active Loads
ReFlex Power™
7-54
M380056-01 Rev L
8. Set the second power supply (Power supply connected to IAI Board)
voltage to 10.00 volts.
9. After the DMM voltage stabilizes enter the voltage of the DMM
(Connected to second power supply or Power supply connected to
IAI Board.)
Note: This voltage should be equal to 10.0 Volts.
CAL<n>:INP:VOLT2 <value-2>
10. Send the command to calculate the gain and offset of the IAI board.
CAL<n>:MEAS:OFFSET:CALC
11. Set the active load back to the SCPI mode
INP<n>:CURRent:EXTernal 0
12. Send the system error query command to make sure calibration was
successful.
SYST:ERR?
13. If there is no error returned at the system error query command then
save the calibration result with the following command
CAL<n>:UNLOCK “6867”
CAL<n>:STORE
CAL<n>:LOCK
14. Reset the unit
*RST<n>
Содержание Elgar ReFlex Power
Страница 1: ...M380056 01 Rev L www programmablepower com ReFlex Power Operation Manual...
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...ii This page intentionally left blank...
Страница 8: ...iv This page intentionally left blank...
Страница 10: ...vi This page intentionally left blank...
Страница 21: ...M380056 01 Rev L xvii This page intentionally left blank...
Страница 22: ......
Страница 85: ...ReFlex Power Controller ReFlex Power 3 22 M380056 01 Rev L This page intentionally left blank...
Страница 89: ...AC Power Supplies ReFlex Power 4 4 M380056 01 Rev L Figure 4 1 ACPS Module Front Panel 875VA...
Страница 120: ...ReFlex Power AC Power Supplies M380056 01 Rev L 4 35 Figure 4 11 Parallel Output Configuration...
Страница 147: ...AC Power Supplies ReFlex Power 4 62 M380056 01 Rev L RST n...
Страница 157: ...AC Power Supplies ReFlex Power 4 72 M380056 01 Rev L 26 Reset the unit RST n...
Страница 173: ...Fixed Power DC Power Supply ReFlex Power 5 4 M380056 01 Rev L Figure 5 2 Rear View of Typical DC Supply Module...
Страница 260: ...ReFlex Power DC Power Supplies M380056 01 Rev L 6 57 Figure 6 22 Series Parallel Output Configuration...
Страница 274: ...ReFlex Power DC Power Supplies M380056 01 Rev L 6 71 18 Turn off the output RST n...
Страница 275: ...DC Power Supplies ReFlex Power 6 72 M380056 01 Rev L This page intentionally left blank...
Страница 279: ...Active Loads ReFlex Power 7 4 M380056 01 Rev L Figure 7 1 LPAL 375W Front Panel...
Страница 280: ...ReFlex Power Active Loads M380056 01 Rev L 7 5 Figure 7 2 HPAL 750W Front Panel...
Страница 281: ...Active Loads ReFlex Power 7 6 M380056 01 Rev L Figure 7 3 Typical Active Load Rear Panel...
Страница 292: ...ReFlex Power Active Loads M380056 01 Rev L 7 17 Figure 7 5 Slot Numbering Assignment Empty Mainframe Front View...
Страница 359: ...Active Loads ReFlex Power 7 84 M380056 01 Rev L This page intentionally left blank...