5-16 | DYCOR LCD RESIDUAL GAS ANALYZER WITH SYSTEM 2000 SOFTWARE
filament Pins 3 and 8. They should be isolated from ground.
−
If no continuity exists between Pins 3 and 8, remove the quadrupole head
from the vacuum system and replace the filament assembly with a replace-
ment filament assembly.
If the problem persists after going through these steps, you will have to localize
the problem area. These potential problem areas are:
• Ethernet LCD Communications and LEDs
• Filament Trip/Overpressure and LEDs
lcd communications and leds
1. Make sure the power supply cable between the Ethernet LCD and the power
Figure 5-7.
Quadrupole head
pinout, view of flange
(atmosphere side).