o
nfiguring the B3000-B
Using the B3000-B
The B3000-B
mistic
serial brain is a drop-in replacement for the
obsolete B3000 serial brain. The B3000-B can be used with:
•
FactoryFloor controllers running OptoControl strategies
•
SNAP PAC S-series controllers, along with other
mistic
I/O units
migrated to PAC Project
•
A PC equipped with an Opto 22 PCI-AC48 adapter card (which
provides an RS-485 port)
NOTE: The B3000-B has the same functionality as the B3000, except
that it
does not support the Optomux protocol
. For Optomux, use an
E1 or E2 brain.
If you are not migrating to the SNAP PAC System but
building a new system, use SNAP PAC SB serial brains instead.
For Help
This technical note covers basic configuration only.
See
form #1781, the
B3000-B User’s Guide
, for complete details. If you
cannot find the answer you need in the user’s guide, please contact
Product Support.
Phone:
800-TEK-OPTO (835-6786)
951-695-3080
Monday through Friday, 7 a.m. to 5 p.m. Pacific
Time
Fax:
951-695-3017
E-mail:
Website:
www.opto22.com
B3000-B Top View
Setting Up Serial Networking
1. Attach an RS-485 serial cable to the serial port. If you are using
mistic
interrupts, wire IRQ+ and IRQ– (pins 6 and 7). See the
B3000-B User’s Guide
for details on network wiring.
2. Rotate the baud rate switch to set the
desired baud rate, as shown in the
table below.
NOTE: Due to timing tolerances, some
baud rates may not work with some
compatible devices.
3. Use the three termination switches to
set termination as follows:
• For the B3000-B at the physical end of
the cable, set all three of the termina-
tion switches to the ON position.
• For all B3000-B units that are not at the
physical end of the cable, set all three of
the termination switches to the OFF
position.
NOTE: Since biasing is normally done at
the controller or computer, the B3000-B
does not include biasing switches.
4. Use the two rotary address switches to
set the unit’s address, as shown on the
following page.
Baud rate
switch
Serial
connector
Termination
switches
Address upper
IRQ LED
Reset switch
TX/RX LED
STAT LED
Pin 1
Address lower
8 9 A
B
C
D
E
F
7
6
5
4
3
2
1
0
Baud Rate Switch
Baud rate
Switch
position
Baud rate
Switch
position
(Reserved)
F
4800 bps
7
230400 bps
E
2400 bps
6
115200 bps
D
1200 bps
5
76800 bps
C
600 bps
4
57600 bps
B
300 bps
3
38400 bps
A
(Reserved)
2
19200 bps
9
(Reserved)
1
9600 bps
8
(Reserved)
0
O
N
1
2
3
IRQ
RX
TX/RX
O
N
1
2
3
IRQ
RX
TX/RX
End of cable
All other positions
Baud rate switch set to A for 38400 bps baud rate. Termination
switches set to “ON” position for 4 wire end of link termination.
Upper address set to 8 and lower address set to 3 for a base address
of 128.
NOTE