Specifications
28
XP2i Operation Manual
POWER
Batteries . . . . . . . . . . . . . . . . . . . . . . . . Three size AA (LR6) batteries.
!
WARNING:
Do not remove or change the batteries in hazardous locations.
Approved Batteries
ATEX/IECEx:
The XP2i is Intrinsically Safe only if powered by one of the following battery types:
Approved Battery Type
Ta=
Marking
Rayovac Max Plus 815
-20 to 50° C
Ex ia IIC T4 Ga
Energizer E91
*
Energizer EN91
*
-20 to 50° C
Ex ia IIC T3 Ga
Duracell MN1500
Replace batteries with approved type in non-hazardous locations only
*
Energizer is manufactured by Energizer Holdings, Inc., and the Eveready Battery Company, Inc.
Many other battery types and models have been tested but failed to meet the requirements for Intrinsic Safety—do not assume other models are equivalent.
The XP2i can be operated from an external power supply (AC adapter kit PN: 2984).
Do not mix battery types or manufacturers.
CSA:
The XP2i is Intrinsically Safe only if powered by one of the following battery types:
CSA:
Le XP2i est un système à sécurité Intrinsèque seulement s’il est alimenté par un des Piles de type suivant:
Approved Battery Type
(Approuvé Type de Batterie)
Ta=
Marking
(Marquage)
Rayovac Max Plus 815
-20 to 45° C
Class I, Division 1, A, B, C, D T4
(Classse I, Division 1, A, B, C, D T4)
Energizer E91
-20 to 50° C
Class I, Division 1, A, B, C, D T3C
(Classse I, Division 1, A, B, C, D T3C)
Energizer EN91
Rayovac Max Plus 815
!
WARNING:
Use of Duracell MN1500 batteries is not approved for CSA certification.
!
WARNING:
Do not use the AC Adapter kit in hazardous locations.
Battery Life . . . . . . . . . . . . . . . . . . . . . . 1500 hours typical (alkaline battery).
Ultra Low Power . . . . . . . . . . . . . . . . . > 1 year, typical with 20 minute recording interval in LT5 mode.
Battery Indicator . . . . . . . . . . . . . . . . . 3-segment Battery Icon: (
)
= Full Battery; (
)
= Used Battery; (
)
= Low Battery
Dead Battery Indication . . . . . . . . . .
batt