Specifications

 

 25

XP2i Operation Manual

Specifications

Accuracy – Gauge Pressure

Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year.

Gauges must be exercised whenever exposed to significant changes in environmental conditions to achieve these specifications, and then re-zeroed. To 
exercise a gauge, cycle the gauge between zero (ambient barometric pressure) and the pressure of interest. A properly exercised gauge will return to a perfect 
zero reading (or return to the same ambient barometric reading).

Exposure to environmental extremes of temperature, shock, and/or vibration may warrant a more frequent recertification period.

0 to 20% of Full Scale . . . . . . . . . . . . ±(0.02% of Full Scale).

20 to 100% of Full Scale . . . . . . . . . . ±(0.1% of Reading).

Vacuum . . . . . . . . . . . . . . . . . . . . . . . . . ±(0.25% of Full Scale), where F.S. = -14.5 psig, -1.0 bar, -100.0 kPa.

Note:

 All models indicate vacuum, but vacuum specification applies only to 300 psi / 20 bar / 2000 kPa and lower pressure gauges.

 

X

Vacuum to Full Scale

-S2 Option:   . . . . . . . . . . . . . . . . . . . . . . ± (0.02% of Full Scale

*

) 15 to 25° C 

± (0.05% of Full Scale

*

) -10 to 15° C, 25 to 50° C, typical

-S5 Option:   . . . . . . . . . . . . . . . . . . . . . . ± (0.05% of Full Scale

*

) -10 to 50° C

Full Scale is the numerical value of the positive pressure range.

 !

CAUTION:  Not recommended for continuous use below -14.5 psig or 0.2 psiA. Refer to the 

XP2i-DP

 data sheet for gauges that are intended for 

continuous high vacuum use.

Содержание Crystal Pressure 200BARXP2I-DD

Страница 1: ...XP2i Operation Manual for XP2i Digital Test Gauge Standard and Dual Display DD...

Страница 2: ...oggerXP Gauge Operation 18 Excel Templates 22 Enclosure 23 Serial Numbers 24 Specifications 25 Power 28 Pressure Ranges Display Scales Resolution 29 Part Numbering System 29 Safety Certifications 30 H...

Страница 3: ...teel sensor Calibration due reminder We hope your XP2i meets your expectations and we re interested in any comments or suggestions you may have You can send us a note at crystal ametek com Many featur...

Страница 4: ...tubing and fittings Be certain all pressure connections are secured Most XP2i s are intended for gauge pressure measurement That is they indicate the dif ference between applied pressure and ambient...

Страница 5: ...g PSV Mode maximum blinking PSV Mode minimum Average pressure DataLoggerXP datalogging mode From the factory this setting is disabled Use ConfigXP to enable Peak High and Peak Low values are not saved...

Страница 6: ...XP defaults to 20 psi the command will be ignored and HI will be displayed X X To Zero the XP2i Turn off peak indication by pressing the peak button repeatedly until the HI and LO icons are off then p...

Страница 7: ...econd line will display the rate and the min icon will display Rate is calculated at every pressure measurement 4 times per second and the displayed value is the average of the most recent 3 to 10 cal...

Страница 8: ...ton again the backlight will go into a lower brightness setting to extend bat tery life and remain on for 2 minutes after the last key is pressed Press the backlight button once more and the light wil...

Страница 9: ...oed when 30 psi is being applied it will indicate that the overpressure condition has been reached at 80 psi i e 110 x 100 psi 30 psi 80 psi Overpressure can affect accuracy but the effect is only tem...

Страница 10: ...y Cal Due Note Protect the settings by using password protection Low Battery Indication The Battery icon uses three bars to display the battery level When the icon displays all three bars the batterie...

Страница 11: ...e I Division 1 A B C D T4 Energizer E91 20 to 50 C Class I Division 1 A B C D T3C Classse I Division 1 A B C D T3C Energizer EN91 Rayovac Max Plus 815 WARNING Use of Duracell MN1500 batteries is not a...

Страница 12: ...ws you to record 100 data points before requiring purchase With an XP2i and DataLoggerXP you can Record up to 32000 pressure measurements data points into non volatile flash memory Save battery life a...

Страница 13: ...Logging Parameters The Logging Interval and Logging Type are referred to in this document as the Logging Parameters Run A set of one or more readings taken by the XP2i The user can specify different...

Страница 14: ...e to download the latest version of the software Once the setup application begins follow the on screen prompts to complete the installation If DataloggerXP is being upgraded the prior version does no...

Страница 15: ...of the application The DataLoggerXP Main Window Working in the Main Window X X Setting the Options By default DataLoggerXP is configured for automatic serial port detection However in some cases auto...

Страница 16: ...e gauge allows you to prevent unauthorized changes to the data logging parameters Once the password is set the Send to Gauge button will require the password Changing logging parameters and Auto Off s...

Страница 17: ...st choose a logging type The gauge may operate in five logging types Actual The gauge stores the value displayed on the gauge at each logging interval Average The gauge calculates and stores the avera...

Страница 18: ...ake up to 15 seconds to download all the data Once the data has been downloaded the View and Save buttons will be available X X To View a Data Run In the DataLoggerXP main window click the View button...

Страница 19: ...dropdown box in the Save As dialog box choose either an Excel template file ending in xlt or an Excel workbook file ending in xls If you choose the Excel template xlt a new Excel document will be crea...

Страница 20: ...splayed To avoid starting a recording release the units button after you see the number of available points The Dual Display DD XP2i will display the number of available points on the second line anyt...

Страница 21: ...allows you to select the logging type and interval from the keypad Press the peak button repeatedly to cycle through the logging types Press the units button to accept the format Actual Average Avera...

Страница 22: ...ange as follows If the gauge is in a Clear or Off state the gauge will briefly display Start begin logging data and change its state to On If the gauge is in an On state it will stop logging briefly d...

Страница 23: ...blink XP2i Reset or Battery Change Data collected by the XP2i will not be lost with a gauge reset or battery change although any run in progress will be stopped A reset or battery change will affect...

Страница 24: ...e dataloggerxp xlt is located in the DataLoggerXP program directory and can be used as is or modified to meet the user s require ments The default template produces a graph with an elapsed seconds tim...

Страница 25: ...PT ADAPTER P N MPM 1 4MPT 1 4 INCH MBSP ADAPTER P N MPM 1 4BSPM 1 30 32 6 M20X1 5 ADAPTER P N MPM M20X1 5M 0 93 23 6 DISTANCE TO DIAPHRAGM INTERNAL PRESSURIZED VOLUME 295 MM3 0 02 IN3 5 59 142 0 3 4 1...

Страница 26: ...under the battery furthest from the serial port connection You may also find your serial number using ConfigXP software See ConfigXP for more information Serial Numbering System Serial Numbers consis...

Страница 27: ...ronmental extremes of temperature shock and or vibration may warrant a more frequent recertification period 0 to 20 of Full Scale 0 02 of Full Scale 20 to 100 of Full Scale 0 1 of Reading Vacuum 0 25...

Страница 28: ...tion Note Calibrated in Vertical Orientation Orientation Effect 0 004 psi X X100 psiA Range 0 to 20 of Range 0 02 of Full Scale 20 to 110 of Range 0 1 of Reading S2 Option 0 02 of Full Scale 15 to 25...

Страница 29: ...U S Patent No 8 794 677 Compatible with HiP LM4 and LF4 Series Autoclave Engr SF250CX Male and Female Series CPF o ring size and material AS568A 012 Viton 90 durometer P N 3981 CAUTION To achieve CPF...

Страница 30: ...ed from an external power supply AC adapter kit PN 2984 Do not mix battery types or manufacturers CSA The XP2i is Intrinsically Safe only if powered by one of the following battery types CSA Le XP2i e...

Страница 31: ...y pressure in kPa MPa and bar or mbar only psi and bar models can display all available units XP2i will indicate pressure up to 10 above Range Pressure Above 110 the XP2i display will flash indicating...

Страница 32: ...aluminium if it is mounted in an area where the use of category 1 G apparatus is required it must be installed such that even in the event of rare incidents ignition sources due to impact and friction...

Страница 33: ...weise F r Explosionsgef hrdeten Orten DEUTSCH German Die RS 232 Schnittstellenverbindung darf niemals in einer explosionsgef hrdeten Umgebung benutzt werden Der Batteriewechsel mu ausschlie lich in si...

Страница 34: ...assume other models are equivalent Energizer is manufactured by Energizer Holdings Inc and the Eveready Battery Company Inc Instrucciones De Seguridad Para Zonas Peligrosas ESPA OL Spanish No use el c...

Страница 35: ...seulement s il est aliment par un des Piles de type suivant Approved Battery Type Ta Marking Rayovac Max Plus 815 20 to 50 C Ex ia IIC T4 Ga Energizer E91 Energizer EN91 20 to 50 C Ex ia IIC T3 Ga Dur...

Страница 36: ...jke Omgeving NEDERLANDS Dutch Het gebruik van de RS232 interface is niet toegestaan in een explosie gevaarlijke omgeving Vervang de batterijen uitsluitend in een niet explosie gevaarlijke omgeving en...

Страница 37: ...s r wnowa ne Energizer jest produkowany przez Energizer Holdings Inc lub przez Eveready Battery Company Inc R j hdysvaarallisten Tilojen Turvallisuusohjeita SUOMEN KIELI Finnish RS 232 v yl liitint EI...

Страница 38: ...ason As in the previous note gas has dissolved into the hydraulic fluid When decreasing the pres sure the dissolved gas then leaves the fluid but at an uneven rate so small pressure differential due t...

Страница 39: ...minimum rated accuracy of 0 025 of reading or equivalent in terms of percent of full scale This level of accuracy requires the use of piston deadweight gauges or very high performance pressure control...

Страница 40: ...will not charge batteries but in the event of AC power loss XP2is will automatically revert to battery operation WARNING Do not use the AC Adapter Kit in hazardous locations P N 3009 Plastic Carrying...

Страница 41: ...ady Battery Company Inc Pressure is Our Business is a registered trademark of Crystal Engineering Corp WARRANTY Crystal Engineering Corporation warrants the XP2i DigitalTest Gauge to be free from defe...

Страница 42: ...4870 Q 2018 Crystal Engineering Corporation 708 Fiero Lane Suite 9 San Luis Obispo California 93401 8701...

Отзывы: