78
ENTRY
DESCRIPTION
BUS
Bus mode. The transient system is started by a bus command
or a group execute trigger (GET).
EXT
External mode. The transient system is started by a user-
provided external TTL trigger signal on TRIGGER IN.
TRAN STEP
Indicates the transient system execution mode. Available modes are:
AUTO
When triggered, the transient system will automatically
execute each list point sequentially without waiting for a
trigger between list points. This execution is paced by the
dwell time set for each data point.
ONCE
When triggered, the transient system will execute the first list
point and wait for a new trigger once the dwell time expires.
This allows triggered execution of each step in the transient
list.
List parameters:
VOLT
Step #
Voltage set point
VSLEW
Step #
Voltage slew rate in V/s
FREQ
Step #
Frequency set point
FSLEW
Step #
Frequency slew rate in Hz/s
DWELL
Step #
Dwell time in seconds.
Range is 0.001 to 900000
TTLT
Step #
ON:
Generates an output trigger pulse at this list step.
OFF:
No output trigger.
The output trigger is available on the TRIG OUT on the rear
panel.
FUNC
Step #
Waveform selection. Available choices are Sinusoid, Square,
Clipped or any of the user provided waveforms in waveform
memory (-1Pi models only).
PHASE
Step #
Phase angle set point. (Not relevant for phase A
if clock mode is internal.)
CURR
Step #
Current set point
Transient List point data entry method.
Transient list points are numbered sequentially from 0 through 99 and executed in this order.
Each list point or list entry has 9 parameters as shown in the table above. To enter list point
data, the keypad must be used. The shuttle knob is used to increment or decrement the list point
sequence number (#). The sequence number can only be increased to the next available empty
(new) list point.
To move to the next or previous parameter, use the UP (
) or DOWN (
) cursor keys
It is not necessary to use all list points, only as many needed to accomplish the desired output
sequence.
Setting Data Values
Data values can be set for each point in a list. If all data values in a specific list are going to be
the same value (e.g. the current limit parameter is set to the same value for the entire transient
program), only the first data value for that parameter has to be set. Setting only the first data
point will automatically repeat that value for all subsequent points in the transient list.
Setting Slew Rates
Содержание California Instruments MX15 Series
Страница 1: ...7005 960 Rev N www programmablepower com MX15 Series AC and DC Power Source User Manual...
Страница 29: ...MX15 Series 29 Figure 2 5 MX output filter option schematic...
Страница 30: ...30 Figure 2 6 MX15 Output Noise 10 KHz 1 MHz...
Страница 31: ...MX15 Series 31 Figure 2 7 MX45 Output Noise 10 KHz 1 MHz with Optional Filter...
Страница 34: ...34 Figure 3 2 Location of AC Input Fuse Block and Chassis Ground Connection Rear View Access Panel Removed...
Страница 36: ...36 Figure 3 3 MX Series AC Input Connection Diagram Rear view...
Страница 43: ...MX15 Series 43 Figure 3 8 Two MX s in Clock and Lock mode Output Wiring Rear view...
Страница 45: ...MX15 Series 45 Figure 3 10 Three MX s in Clock and Lock mode Output Wiring Rear view...
Страница 58: ...58 Figure 3 15 Functional Test Setup...
Страница 106: ...106 Figure 5 2 MX Series Detailed Block Diagram...
Страница 107: ...MX15 Series 107 Figure 5 3 Power Module Detailed Block Diagram...
Страница 111: ...MX15 111 Figure 5 5 Amplifier Board Layout...
Страница 139: ...MX15 139 Figure 9 4 Voltage Modulation Frequency characteristics...