6.3.1 Measurement Cal
AC Volt Full-scale:
Program the output to the 300 VAC range. Close the output
relay. Program the output to 240 VAC and 60 Hz. Go to the
MEASUREMENT CALIBRATION screen. Enter the actual AC
output voltage for the VOLT FS parameter and press the
ENTER key.
Save this value by pressing the SET key.
Note:
This process may take as long as a few
minutes
to
complete after the enter key is pressed.
AC Current Full-scale:
Apply a load to the output. Refer to Table 6-1. Program the
output to 120 VAC on the 150 VAC range and 60 Hz.
Observe the actual output current and enter this value for the
CURR FS parameter. Press the ENTER key.
Save this value
by pressing the SET key.
DC Volt Full-scale:
Program the output to the 400 VDC range. Close the output
relay. Program the output to 320 VDC. Go to the
MEASUREMENT CALIBRATION screen. Enter the actual AC
output voltage for the VOLT FS parameter and press the
ENTER key.
Save this value by pressing the SET key.
Note:
This process may take as long as a few
minutes
to
complete after the enter key is pressed.
DC Current Full-scale:
Apply a load to the output. Refer to Table 6-1. Program the
output to 160 VDC on the 200 VDC range.
Observe the actual output current and enter this value for the
CURR FS parameter. Press the ENTER key.
Save this value
by pressing the SET key.
6.3.2 Measurement Calibration Summary
The following Table is a summary of the preceding calibration steps. The value indicated by the
External DVM is called V
AC
or V
DC
. The current measured by the current shunt is called I
AC
or
I
DC
.
TITLE
PROGRAM/LOAD PARAMETERS
PARAMETER
ADJUST TO
AC MODE
AC Volt Full-scale
300 VAC Range, 240 VAC, 60 Hz, no load
VOLT FS
V
AC
AC Current Full-
scale
150 VAC Range, 120 VAC, 60 Hz, full load to
90% of max current range.
CURR FS
I
AC
DC MODE
DC Volt Full-scale
400 VDC Range, 320 VDC, no load
VOLT FS
V
AC
DC Current Full-
scale
200 VDC Range, 160 VDC, full load to 90% of
max current range.
CURR FS
I
AC
Table 6-2: Measurement Calibration Table
Содержание California Instruments MX15 Series
Страница 1: ...7005 960 Rev N www programmablepower com MX15 Series AC and DC Power Source User Manual...
Страница 29: ...MX15 Series 29 Figure 2 5 MX output filter option schematic...
Страница 30: ...30 Figure 2 6 MX15 Output Noise 10 KHz 1 MHz...
Страница 31: ...MX15 Series 31 Figure 2 7 MX45 Output Noise 10 KHz 1 MHz with Optional Filter...
Страница 34: ...34 Figure 3 2 Location of AC Input Fuse Block and Chassis Ground Connection Rear View Access Panel Removed...
Страница 36: ...36 Figure 3 3 MX Series AC Input Connection Diagram Rear view...
Страница 43: ...MX15 Series 43 Figure 3 8 Two MX s in Clock and Lock mode Output Wiring Rear view...
Страница 45: ...MX15 Series 45 Figure 3 10 Three MX s in Clock and Lock mode Output Wiring Rear view...
Страница 58: ...58 Figure 3 15 Functional Test Setup...
Страница 106: ...106 Figure 5 2 MX Series Detailed Block Diagram...
Страница 107: ...MX15 Series 107 Figure 5 3 Power Module Detailed Block Diagram...
Страница 111: ...MX15 111 Figure 5 5 Amplifier Board Layout...
Страница 139: ...MX15 139 Figure 9 4 Voltage Modulation Frequency characteristics...