background image

LSP

HD

 System

LSP

HD

 Linescanner Head

User Guide

Publication Nº 805233

Issue 7

Language: English

25 May 2016

© Land Instruments International, 2016

Содержание 0 LAND LSP HD 61

Страница 1: ...LSPHD System LSPHD Linescanner Head User Guide Publication N 805233 Issue 7 Language English 25 May 2016 Land Instruments International 2016...

Страница 2: ...dous Read and understand the user documentation supplied before installing and operating the equipment Protective Clothing Face and Eye Protection It is possible that this equipment is to be installed...

Страница 3: ...e problem together with your own name and contact information address telephone number email address etc Design and Manufacturing Standards The Quality Management System of Land Instruments Internatio...

Страница 4: ...Blank...

Страница 5: ...Emissivity Typical Values 21 5 1 Setting the Emissivity value 21 5 2 LSPHD 1 Series 23 5 4 LSPHD 5 Series 23 5 3 LSPHD 2 Series 23 5 5 LSPHD 6 Series 24 5 6 LSPHD 7 Series 24 6 Setting the Scan Freque...

Страница 6: ...PHD 50 LSPHD 61 LSPHD 20 LSPHD 52 LSPHD 62 LSPHD 21 LSPHD 22 The User Guide covers the installation operation and maintenance of the LSPHD Linescanning Thermometer The instructions given in this User...

Страница 7: ...lectronics to provide accurate linescan temperature measurement at high scan rates The system is compact yet rugged providing reliable operation in the most severe industrial environments The LANDSCAN...

Страница 8: ...s LSP HD 6 System Name Series Number Temperature Range 1 1 Nomenclature The LSP Linescanning Thermometer detail label can be found on the black scanner body section The label displays the scanner seri...

Страница 9: ...ng wavelength 1 m 1 m Temperature range 600 to 1400 C 1112 to 2552 F 700 to 1500 C 1292 to 2732 F System Accuracy 1 5 to 95 of range 2 C 3 6 F Focus distance 2 1m 39 7in to infinity continuously adjus...

Страница 10: ...ians 500 1 0 002 radians Repeatability 0 5 C 0 9 F Temperature Resolution 3 5 of range noise 15 of range 30 of range Above 50 of range 6 C 10 8 F 6 C 10 8 F 2 C 3 6 F 1 C 1 8 F 6 C 10 8 F 6 C 10 8 F 4...

Страница 11: ...2 Below 15 of range noise Above 15 of range 1 5 C 2 7 F 1 C 1 8 F 1 5 C 2 7 F 1 C 1 8 F 2 5 C 4 5 F 2 5 C 4 5 F Scan Angle 3 80 Scan Rate Factory set 20 Hz 50 Hz 50 Hz User adjustable 10 to 150Hz in...

Страница 12: ...Below 15 of range 1 5 C 2 7 F 1 5 C 2 7 F 2 5 C 4 5 F 2 C 3 6 F noise Above 15 of range 1 C 1 8 F 1 C 1 8 F 2 5 C 4 5 F Scan Angle 3 80 Scan Rate 10 to 150Hz in 1 Hz steps Factory set 20 Hz 50 Hz 50...

Страница 13: ...C 0 9 F Temperature Resolution 2 Below 15 of range noise Above 15 of range 4 C 7 2 F 2 C 3 6 F Scan Angle 3 80 Scan Rate Factory set 50 Hz User adjustable 10 to 150 Hz in 1 Hz steps Speed of response...

Страница 14: ...d fumes are kept to a minimum Ensure that the scanning sight path is not obscured The LSPHD 10 20 21 have a focus distance factory set at 2000 mm 79 in Refer to Section 2 2 for details of how to chang...

Страница 15: ...quired to match the focus distance you want to set The table below gives the focus distances that can be achieved by turning the focusing ring Each screw hole on the focusing ring is at 0 05 of a turn...

Страница 16: ...bles for x and y axis sizes at 40 2 3 Positioning The Scanner Before installing the scanner refer to the cautions given in Section 2 1 The scanner must be installed so that the target plane is paralle...

Страница 17: ...40 the length of the Target Path TP between the scanner and the target is TP TD 35 0 766 mm or TP TD 1 38 0 766 inches Note The active scan angle may be set in the range 40 to 80 See the Landscan Conf...

Страница 18: ...0 39 1750 68 9 2996 118 0 2284 89 9 3 5 0 14 10 5 0 41 8 0 0 31 5 8 0 23 14 4 0 57 11 0 0 43 2000 78 7 3415 134 4 2611 102 8 4 0 0 16 11 4 0 45 8 7 0 34 6 7 0 26 16 0 0 63 12 2 0 48 2500 98 4 4254 16...

Страница 19: ...axis mm in mm in mm in mm in mm in mm in 200 7 9 394 15 5 261 10 3 12 0 0 47 15 7 0 62 12 0 0 47 500 19 7 898 35 4 653 25 7 12 0 0 47 15 7 0 62 12 0 0 47 750 29 5 1317 51 9 979 38 5 12 0 0 47 15 7 0 6...

Страница 20: ...n how to use the laser alignment unit refer to Section 4 0 of this User Guide To interpret the scan profile produced by the scanner the scan direction of the unit must be known The scan direction is i...

Страница 21: ...nter the scanner 2 6 3 Compensating for losses in viewing windows The emissivity control on the LSPHD can be set to compensate for target emissivity and window losses such as when using the LSP Sealed...

Страница 22: ...d Scanner and is a right angled M12 D coded Ethernet connector The power supply end of the cable is a straight M12 D coded Ethernet connector 3 1 1 Power supply units 807105 LSPHD Power Over Ethernet...

Страница 23: ...ions G1 8 tappings in the plate edge There is no preferred flow orientation However if the scanner is mounted in a position where one of the water connectors is higher than the other then the lower co...

Страница 24: ...path Connect the air supply via the G3 8 threaded hole at the end of the mounting plate If possible ensure that the air purge exit direction is downwards Make sure that air purge exit is not impeded b...

Страница 25: ...tion where a direct mirror like reflection may enter the eye WARNING CLASS 2 Laser Product DO NOT attempt to disassemble the laser unit or any of its mounting components Embedded laser has a maximum o...

Страница 26: ...ripe of laser light is projected through the window of the scanner indicating the position and extents of the scan line 2 Adjust the scanner position as required so that the product almost fills the w...

Страница 27: ...lternatively the emissivity setting can be set to a fixed value using the procedure below 1 Unscrew the four bolts holding the scanner side plates to the body using a 5mm allen hex key 2 Notice that t...

Страница 28: ...osed part of the second layer of PCB Replace the jumper removed in step 3 on the TOP TWO of the three pins of LK3 marked Internal Emissivity 5 The scanners emissivity can now be set using the tow iden...

Страница 29: ...85 Palladium 0 28 Platinum 0 27 Rhodium 0 25 Silver 0 05 oxidised 0 10 Tin 0 40 oxidised 0 60 Titanium 0 55 oxidised 0 80 Tungsten 0 39 Zinc 0 50 oxidised 0 60 Metals Emissivity Aluminium 0 09 oxidis...

Страница 30: ...idised 0 45 Nichrome oxidised 0 90 Metals Emissivity Aluminium oxidised 0 30 Copper oxidised 0 85 Iron Steel oxidised 0 85 Lead oxidised 0 63 Molybdenum oxidised 0 80 Titanium 0 30 Miscellaneous Emiss...

Страница 31: ...ls are required 5mm hexagon key 3mm flat blade screwdriver plastic 1 Remove the scanner interface plug This ensures that the scanner unit is electrically isolated 2 When viewed from the top of the sca...

Страница 32: ...d 6 2 Scan Frequency Table Switch Position Nominal Scan Frequency Hz 1 20 2 40 3 60 4 80 5 100 6 120 7 120 8 150 9 N A 0 N A M6 cap head screws Scan direction arrow End plate End plate seal Scan frequ...

Страница 33: ...dely spaced Lateral resolution Lower resolution Higher resolution Disk space for file storage Greater space required Less space required Processor requirements Higher Lower 6 4 Data Storage Requiremen...

Страница 34: ...mpared to Example 1 The scan skew angle is given by e g At a scan frequency of 50Hz a target distance D of 2m and product speed of 10m s 0 91 Apparent measured product width actual product width ignor...

Страница 35: ...et to 50 C 122 F This gives an indication of possible over temperature within the instrument The maximum operating limit for the LSPHD Linescanner Head is 60 C 140 F If the indicated internal temperat...

Страница 36: ...the purge system is of high quality and is supplied clean and dry otherwise window contamination may result and the diffuser within the purge unit may become blocked If the viewing window becomes dirt...

Страница 37: ...AMETEK Land to aid installation of the LSPHD Linescanning Thermometer and to ensure that the equipment is compatible with the intended application environment For detailed information on installation...

Страница 38: ...de gives you instructions on how to read the version information on WCA LSP LSP HD and IO devices 10 1Reading the WCA program version 1 In the Landscan WCA software open the Help menu and select the A...

Страница 39: ...ng the LSP HD Scanner version information 1 In the Landscan WCA software open the Diagnostics menu then select LPU LSC The LPU LSC Diagnostics window is displayed 2 Make a note of the Scanner Version...

Страница 40: ...version 5 9 read the version from the label printed on the IO unit As shown below For users with WCA version 5 10 0 or later go to step 2 2 For users of WCA version 5 10 0 or later select the Configur...

Страница 41: ...truments based on System 4 UNO thermometers accessories and mountings and special instruments based on UNO Application dedicated processors based on LANDMARK Graphic ABTS S and ABTS U FTS VDT S and VD...

Страница 42: ...ions outside the Ambient Temperature Humidity and Vibration rating of the product environmental contamination solvent vapours deposition of airborne contamination cooling liquids of non neutral pH etc...