Test voltage drop in diodes�
WarninG!
To prevent electric shock, turn
off all power and fully discharge capacitors
on the circuit under test before measuring�
1� Plug black test lead into
cOM
Jack�
Plug red test lead into
input
Jack�
2� Turn the Rotary Switch to the
Ω
position�
3� Press
Func
until
is displayed�
4� Connect red probe to diode’s anode
and black probe to its cathode�
5� Read measured forward-biased
voltage drop on the Display�
note:
If the test leads are reversed
or open,
OL
is displayed�
6� When testing is complete, turn
Rotary Switch to
OFF
, remove
Test Leads and store with Meter�
capacitance Measurements
Measure capacitance up to 99�99mF�
WarninG!
Turn off all power and fully
discharge capacitors on the circuit
under test before measuring�
1� Plug black test lead into
cOM
Jack�
Plug red test lead into
input
Jack�
2� Turn the Rotary Switch
to the position�
3� Carefully touch capacitor
leads with tips of probes�
4� Read measured capacitance
on the Display�
5� When testing is complete, turn
Rotary Switch to
OFF
, remove
and store capacitor and Meter�