Pump P-960
User Manual 18-1172-73 Edition AA
Measured volume too low
Air bubbles are trapped in the
pump
Purge the pump according to
section 5.2
Cavitation might occur due to
too high flow rate for the
sample application technique
used
See the ÄKTAdesign Optional
Configurations User Manual
for the recommended flow
rates
A check valve is clogged or
damaged, causing leakage
Remove the check valve
according to section 4.6.2.
Clean the check valve in an
ultrasonic bath. If the leakage
persists, replace the
assembly.
The flow rate is not calibrated
properly
Contact Amersham
Biosciences for advice
Erratic flow or pressure pulsation
Leaking tubing connector
Check the tubing connectors.
Tighten or replace if
necessary.
The solvent filter is clogged
Check the solvent filter. Clean
if necessary.
The pump inlet or outlet
tubing is pinched or damaged
Check the pump tubing.
Replace if necessary.
Air bubbles are trapped in the
pump
Purge the pump according to
section 5.2
A check valve is clogged or
damaged, causing leakage
Remove the check valve
according to section 4.6.2.
Clean the check valve in an
ultrasonic bath. If the leakage
persists, replace the
assembly.
Error symptom
Possible cause
Corrective action
18117273aa.book Page 26 Thursday, July 3, 2003 5:13 PM
Содержание AKTA design P-960
Страница 1: ...Pump P 960 18 1172 73 User Manual 18117273aa book Page 1 Thursday July 3 2003 5 13 PM...
Страница 2: ...18117273aa book Page 2 Thursday July 3 2003 5 13 PM...
Страница 4: ...18117273aa book Page 4 Thursday July 3 2003 5 13 PM...
Страница 39: ...18117273aa book Page 39 Thursday July 3 2003 5 13 PM...