Troubleshooting
5
Pump P-960
User Manual 18-1172-73 Edition AA
25
5 Troubleshooting
5.1
Error symptoms and corrective actions
If the suggested actions do not correct the fault, contact your local
service organisation.
WARNING! LETHAL VOLTAGE.
Always disconnect the power supply
before attempting to replace any item on the unit.
WARNING! HAZARDOUS CHEMICALS.
Incorrectly fitted tubing
might loosen, causing a jet of liquid to spray out. This is especially
dangerous if hazardous chemicals are being used. Connect the
tubing by first inserting the tubing fully, then tightening the connector
fingertight.
CAUTION
! Only spare parts approved or supplied by Amersham
Biosciences may be used for maintaining and servicing the pump.
Error symptom
Possible cause
Corrective action
Leakage
A tubing connector is
tightened poorly, or damaged
Check all tubing connectors
for leakage. Tighten or replace
if necessary.
The pump inlet or outlet
tubings are pinched or
damaged
Check the pump tubing.
Replace if necessary.
A check valve is clogged or
damaged
Remove the check valve
according to section 4.6.2.
Clean the check valve in an
ultrasonic bath. If the leakage
persists, replace the check
valve.
An O-ring in a check valve or
in the connection part is
damaged
Examine the O-rings. If
necessary, replace them
according to the sections 4.5
or 4.6.4.
18117273aa.book Page 25 Thursday, July 3, 2003 5:13 PM
Содержание AKTA design P-960
Страница 1: ...Pump P 960 18 1172 73 User Manual 18117273aa book Page 1 Thursday July 3 2003 5 13 PM...
Страница 2: ...18117273aa book Page 2 Thursday July 3 2003 5 13 PM...
Страница 4: ...18117273aa book Page 4 Thursday July 3 2003 5 13 PM...
Страница 39: ...18117273aa book Page 39 Thursday July 3 2003 5 13 PM...