background image

 

 2  

98-2024 Rev. A 

 

ENDOTOXIN FILTER: 
 

The Endotoxin-Retentive Filters are the final purification process prior to distribution of the 
product water.  They are sub-micron filters with the ability to remove bacteria and endotoxin.  
The filters are enclosed in an opaque housing to inhibit the proliferation of algae.  Water flows 
through the filter, which traps bacteria and endotoxin. 
 
The RAPID REPONSE WATER SYSTEM is shipped completely assembled with required 
water treatment components.  This Operation Manual was written specifically for the RAPID 
REPONSE WATER SYSTEM.  Your system was thoroughly tested and in excellent condition 
when it was shipped to you.  However, because damage during shipment is possible, please 
unpack and carefully inspect the system as soon as you receive it.  Please notify AmeriWater 
if any problems are encountered. 
 

Please read the Operations Manual before using the system.

  Contact AmeriWater 

Customer Service with any questions at 1-800-535-5585 Monday through Friday 8:00 a.m. to 
5:00 p.m. eastern standard time.  For after hours emergencies call 1-800-535-5585 and 
follow the instructions on the recorded message.  Our on-call technician will return your call 
as soon as possible.  This entire Operations Manual should be read before operating or 
servicing the system.  This Operations Manual should then be kept near the system and used 
as a reference and troubleshooting guide. 

 
 1.1     SPECIFICATIONS 

 

Connections 

 

SRO Feed = 1/2" O.D. Hose 

SRO Product  = 1/2" O.D. Hose 

SRO Reject to Drain = 1/2" O.D. Tube 

 

Wall Mount DI Feed = 1/2

” O.D. Tube 

Wall Mount DI Product = 1/2

” MPT to 3/4” FGHT 

 

Storage Tank Feed = 3/4” Male GHT 

Storage Tank Product = 3/4” FPT 

Storage Tank Return = 3/4” FPT 

 

Endotoxin Filter Feed = 3/4”FPT 

En

dotoxin Product = 3/4”FPT 

 

 

Electrical Requirements 

 

 

3X 120V 1PH Dedicated 20Amp GFI service 

 

Shipping Weight 

Operating Weight 

353 LBS 
796 LBS 

 
   Note: Water with a silt density index (SDI) above 5 SDI will foul the membrane. 

 
 

Содержание RAPID RESPONSE

Страница 1: ...98 2024 Rev A RAPID RESPONSE WATER SYSTEM FOR STERIS OPERATION MAINTENANCE MANUAL Manufactured With Pride In The USA www ameriwater com 800 535 5585 AmeriWater 3345 Stop 8 Road Dayton OH 45414...

Страница 2: ...98 2024 Rev A...

Страница 3: ...YSTEM INSTALLATION 5 2 1 SRO 5 2 2 WALL MOUNT DI 6 2 3 STORAGE TANK 7 2 4 ENDOTOXIN FILTER 8 3 0 SYSTEM START UP 9 3 1 OPERATION 9 3 2 BYPASS PROCEDURE 10 4 0 MONITORING 11 5 0 PREVENTATIVE MAINTENANC...

Страница 4: ...cation resin bead then an anion then another cation another anion and so on An exchange takes place of course only when a positive ion contacts a negative exchange site and vice versa With each excha...

Страница 5: ...ns Manual before using the system Contact AmeriWater Customer Service with any questions at 1 800 535 5585 Monday through Friday 8 00 a m to 5 00 p m eastern standard time For after hours emergencies...

Страница 6: ...ures and voltage draws please refer to listed component manuals as listed below For 0021003 Loop Endotoxin Filter refer to Manual 98 0033 For 00SRO1095 RO refer to Manual 98 0150 For 00MW200 Silex Wal...

Страница 7: ...4 98 2024 Rev A 1 2 SYSTEM FLOW DIAGRAM Figure 1 1...

Страница 8: ...e wall with anchoring bolts 3 SRO is equipped with sediment and chlorine removal 4 Connect the feed water supply to the inlet 1 2 fitting on the SRO prefilter 5 The feed water supply requires a minimu...

Страница 9: ...ght of the SRO just before the storage tank and within 6 feet from a 120 volt 60 hertz receptacle 2 Assembly should be level when mounted to the wall 3 Connect tubing from the SRO product flow meter o...

Страница 10: ...ansformer into the receptacle and open the inlet and sample port valves 2 3 STORAGE TANK 1 Locate the storage tank on a firm level foundation and anchored securely 2 Connect the Wall Mount DI product...

Страница 11: ...torage tank to the SRO controller and lock in place Figure 2 4 2 4 ENDOTOXIN FILTER 1 Locate and install the Endotoxin Retentive Filter assembly on or near the wall 2 Connect the 3 4 pump discharge ho...

Страница 12: ...8 0160 For 00HC 0090 PRV Healthcare Storage Tank refer to Manual 098 0001 For 0021003 Loop Endotoxin Filter refer to Manual 98 0033 3 1 OPERATION Once the system has been started the SRO will continue...

Страница 13: ...ce your system components Important note If you have a return flow line coming back to the storage tank when using the bypass the returning water from your process will overflow the tank as the distri...

Страница 14: ...uality management system using continuous reading monitors that measure product water quality total dissolved solids These measurements can provide a measure of equipment performance 3 Cartridge filte...

Страница 15: ...PARAMETERS INITIALS Feed Water Temperature 50 F 80 F Feed Water Pressure 20PSI Pump PSI 100 200PSI Product Flow GPM 0 55 0 76 Reject Flow GPM ADJUSTED EQUAL TO PRODUCT Product Quality TDS 25ppm Total...

Страница 16: ...r 20 6001 must be replaced in the event of listed instances below 1 Red light the red light warns that the water conductivity is above the threshold value of 200 000 ohm For 0185 8000 Storage Tank Ven...

Страница 17: ...ees F Gages are located on face plate of the RO and top of the Pre filter housing Return on the Storage Tank and the inlet and outlet of Endotoxin filter Fig 5 1 Step 4 Membranes Remove membranes from...

Страница 18: ...15 98 2024 Rev A Figure 5 1...

Страница 19: ...END CAP 2 25 and 40 SS HSG SMALL BLUE O RING RO Membrane 21 0035 O RING FLTR HOUS 10 SLIMLINE BLUE BLK 50I O NO PR Prefilter Vent Filter and DI 21530233 O RING FLTR HSG 10IN BB SQ THD Loop Endotoxin...

Отзывы: