background image

 

 1  

98-2024 Rev. A 

1.0     THEORY OF OPERATION 

 
 
This system is designed to pretreat and purify water for use in health care applications. Below 
is the sequence of the water progression as it goes thru the components of the MOBILE SPD 
SYSTEM. Each component’s theory of operation is listed in order of water progression thru 
the system. 
 

RO: 
 

The process of osmosis can be reversed by placing pressure upon the feed water side 
(concentrated solution side) of the membrane.  Water will be forced through the membrane 
barrier to yield water that is purer on the lower pressure side of the membrane than on the 
more concentrated solution side (higher pressure side) of the membrane.  The feed water will 
become more “concentrated,” and will be discharged through the reject port known as “reject 
water” or “concentrate”.  Liberation of purer water from its solutions is caused by the reversal 
of the osmotic pressure; (the operation termed as “Reverse Osmosis”).  Reverse Osmosis is 
commonly referred to as “RO”. 
 

DI: 
 

Mixed Bed De-ionizer contains both the cation and anion resins in a single vessel. The 
purpose of mixing the resins is to achieve a very high quality deionized water. As water 
passes through the mixed bed, it has millions of chances to contact a cation resin bead, then 
an anion, then another cation, another anion, and so on. An exchange takes place, of course, 
only when a positive ion contacts a negative exchange site, and vice versa. With each 
exchange, purity of the water improves because more ions are removed and held by resin 
beads. Mixed bed deionizer is utilized as a polising unit downstream of the RO. 
 

STORAGE TANK: 
 

The reverse osmosis unit makes a limited amount of water at a time. It is therefore necessary 
to store it in a storage tank. The storage tank is equipped with float switches, which will turn 
the reverse osmosis unit off and on when the water reaches a predetermined point in the 
storage tank.  
 
This tank is also equipped with a spray nozzle on the return loop side of the tank. This allows 
disinfectant to reach all parts of the tank when it is applied to the tank, for faster disinfections.  
 
The vent filter on the tank is designed to filter the air when water moves in and out of the 
storage tank. Tanks are equipped with a “rupture disk”, which will break if the flow of air/water 
becomes too much for the filter or tank to handle.

 

The storage tank is equipped with 1 float switches, which will turn the  
reverse osmosis unit off when the water reaches the predetermined cutoff point and 

turn the RO on when the tank is drained to the appropriate level float.  The 
distribution pump on the storage tank will cycle whenever there is a pressure drop 
detected.  

 

Содержание RAPID RESPONSE

Страница 1: ...98 2024 Rev A RAPID RESPONSE WATER SYSTEM FOR STERIS OPERATION MAINTENANCE MANUAL Manufactured With Pride In The USA www ameriwater com 800 535 5585 AmeriWater 3345 Stop 8 Road Dayton OH 45414...

Страница 2: ...98 2024 Rev A...

Страница 3: ...YSTEM INSTALLATION 5 2 1 SRO 5 2 2 WALL MOUNT DI 6 2 3 STORAGE TANK 7 2 4 ENDOTOXIN FILTER 8 3 0 SYSTEM START UP 9 3 1 OPERATION 9 3 2 BYPASS PROCEDURE 10 4 0 MONITORING 11 5 0 PREVENTATIVE MAINTENANC...

Страница 4: ...cation resin bead then an anion then another cation another anion and so on An exchange takes place of course only when a positive ion contacts a negative exchange site and vice versa With each excha...

Страница 5: ...ns Manual before using the system Contact AmeriWater Customer Service with any questions at 1 800 535 5585 Monday through Friday 8 00 a m to 5 00 p m eastern standard time For after hours emergencies...

Страница 6: ...ures and voltage draws please refer to listed component manuals as listed below For 0021003 Loop Endotoxin Filter refer to Manual 98 0033 For 00SRO1095 RO refer to Manual 98 0150 For 00MW200 Silex Wal...

Страница 7: ...4 98 2024 Rev A 1 2 SYSTEM FLOW DIAGRAM Figure 1 1...

Страница 8: ...e wall with anchoring bolts 3 SRO is equipped with sediment and chlorine removal 4 Connect the feed water supply to the inlet 1 2 fitting on the SRO prefilter 5 The feed water supply requires a minimu...

Страница 9: ...ght of the SRO just before the storage tank and within 6 feet from a 120 volt 60 hertz receptacle 2 Assembly should be level when mounted to the wall 3 Connect tubing from the SRO product flow meter o...

Страница 10: ...ansformer into the receptacle and open the inlet and sample port valves 2 3 STORAGE TANK 1 Locate the storage tank on a firm level foundation and anchored securely 2 Connect the Wall Mount DI product...

Страница 11: ...torage tank to the SRO controller and lock in place Figure 2 4 2 4 ENDOTOXIN FILTER 1 Locate and install the Endotoxin Retentive Filter assembly on or near the wall 2 Connect the 3 4 pump discharge ho...

Страница 12: ...8 0160 For 00HC 0090 PRV Healthcare Storage Tank refer to Manual 098 0001 For 0021003 Loop Endotoxin Filter refer to Manual 98 0033 3 1 OPERATION Once the system has been started the SRO will continue...

Страница 13: ...ce your system components Important note If you have a return flow line coming back to the storage tank when using the bypass the returning water from your process will overflow the tank as the distri...

Страница 14: ...uality management system using continuous reading monitors that measure product water quality total dissolved solids These measurements can provide a measure of equipment performance 3 Cartridge filte...

Страница 15: ...PARAMETERS INITIALS Feed Water Temperature 50 F 80 F Feed Water Pressure 20PSI Pump PSI 100 200PSI Product Flow GPM 0 55 0 76 Reject Flow GPM ADJUSTED EQUAL TO PRODUCT Product Quality TDS 25ppm Total...

Страница 16: ...r 20 6001 must be replaced in the event of listed instances below 1 Red light the red light warns that the water conductivity is above the threshold value of 200 000 ohm For 0185 8000 Storage Tank Ven...

Страница 17: ...ees F Gages are located on face plate of the RO and top of the Pre filter housing Return on the Storage Tank and the inlet and outlet of Endotoxin filter Fig 5 1 Step 4 Membranes Remove membranes from...

Страница 18: ...15 98 2024 Rev A Figure 5 1...

Страница 19: ...END CAP 2 25 and 40 SS HSG SMALL BLUE O RING RO Membrane 21 0035 O RING FLTR HOUS 10 SLIMLINE BLUE BLK 50I O NO PR Prefilter Vent Filter and DI 21530233 O RING FLTR HSG 10IN BB SQ THD Loop Endotoxin...

Отзывы: