MRO2X
– 5X Manual
- 42 -
98-0058
Rev. R
PROBLEM
POSSIBLE CAUSE
CORRECTIVE ACTION
Low product flow rate
Low pressure feeding
membrane
Low pump PSI
Reject GPM flow rate too
high
Excessive PRODUCT line
backpressure
Low temperature incoming
tap water
Prefilter clogged
Membrane needs replaced
Verify that the incoming tap
water supply is fully open.
The pressure on the prefilter
gauges should be > 20 PSI
when the MRO is operating.
Pump should be operating at
175
– 250 PSI.
Adjust Reject Water flow rate.
Check for restrictions in the
PRODUCT WATER FOR
DIALYSIS hose. Check the
feed pressure gauge on the
dialysis machine. If the
dialysis machine does not
have a pressure gauge,
install one inline.
Consult the Temperature
Correction Chart to determine
if the flow rate is normal in
relation to the feed water
temperature.
Check the prefilter gauges for
pressure drop. Replace the
prefilter cartridge if the
pressure drop is > 10 PSI.
Replace the membrane.
High Bacteria Count
Too long since the last
disinfection or the procedure
was not performed correctly
Disinfect the MRO following
the procedures in Section 5.
Содержание MRO2X
Страница 2: ......
Страница 9: ...MRO2X 5X Manual 5 98 0058 Rev R SECTION 3 COMPONENTS AND SCHEMATICS 3 1 EXTERNAL FRONT VIEW ...
Страница 11: ...MRO2X 5X Manual 7 98 0058 Rev R 3 2 INTERNAL REAR VIEW ...
Страница 13: ...MRO2X 5X Manual 9 98 0058 Rev R 3 3 ELECTRICAL DIAGRAM MRO2X MRO3X MRO4X MRO5X ...
Страница 14: ...MRO2X 5X Manual 10 98 0058 Rev R 3 4FLUID DIAGRAM MRO2X MRO3X MRO4X ...
Страница 15: ...MRO2X 5X Manual 11 98 0058 Rev R 3 5 FLUID DIAGRAM MRO5X ...
Страница 35: ...MRO2X 5X Manual 31 98 0058 Rev R FIGURE 6 2 ...
Страница 36: ...MRO2X 5X Manual 32 98 0058 Rev R FIGURE 6 3 ...
Страница 52: ...MRO2X 5X Manual 48 98 0058 Rev R SECTION 11 SPARE PARTS LISTING ...
Страница 53: ...MRO2X 5X Manual 49 98 0058 Rev R ...