- 13 -
98-0058
Rev. R
verify that it is within the specified range (pH Water/Bicarbonate/Dialysate Test Strips P/N
97PH20901).
CAUTION:
Mixing chlorine and hydrogen peroxide/peroxyacetic acid causes a toxic
chemical reaction. Never allow them to mix!
Do not use chlorine to disinfect
the system!
8. Use only the exact amount of hydrogen peroxide/peroxyacetic acid disinfectant solution
(PAA) and in proper dilution during disinfection of the system.
9. It is important to test for PAA in the Product Water after rinsing during disinfection of the
system. Do not use the system until all traces of the disinfecting solution in the Product
Water are gone.
CAUTION:
Never operate the MRO connected to dialysis machines if the water
conductivity exceeds the set point, indicating Poor Quality!
10. Always maintain water flow and pressure to avoid damage to the pump.
11. Minimum feed pressure is 20 PSI (while the MRO is in operation, with flow). Maximum feed
pressure is 90 PSI.
12. If the system is operated without a micron prefilter, the membranes will foul.
WARNING: The Clean In Place Switch, located inside the cabinet on the back of the
controller, must be in the OFF position during normal operation. If the
Clean In Place Switch is left in the ON position during normal operation, all
MRO fail-safe modes will be disabled, and damage to the MRO or injury to
the patient may occur.
13.
Minimize the opportunities for bacterial growth between uses!
Whenever the MRO is not used for a period of several hours, and connected to a Direct
Feed Loop, the “Membrane Flush Feature” of the MRO should be programmed to be
active when in the STANDBY mode (See section 5 for activating this feature).
Whenever the MRO is not used for a period of several hours, and connected to a
Storage Tank of a Central System, the “Membrane Flush Feature” of the MRO should be
programmed to be active when in the OPERATE mode (See section 5 for activating this
feature). This feature will flush the MRO when the Storage Tank is full and the water
level is maintained at the Tank Full High float switch by diverting the Product Water to
drain.
Before determining a bacteria count, the MRO should be PLACED in FLUSH for 5
minutes (fully opened reject flow control knob), and then placed in OPERATE for 5
– 10
additional minutes after a period of non-use, but before taking a sample of the
PRODUCT WATER. Bacteria are known to increase in population when water is not
moving.
Содержание MRO2X
Страница 2: ......
Страница 9: ...MRO2X 5X Manual 5 98 0058 Rev R SECTION 3 COMPONENTS AND SCHEMATICS 3 1 EXTERNAL FRONT VIEW ...
Страница 11: ...MRO2X 5X Manual 7 98 0058 Rev R 3 2 INTERNAL REAR VIEW ...
Страница 13: ...MRO2X 5X Manual 9 98 0058 Rev R 3 3 ELECTRICAL DIAGRAM MRO2X MRO3X MRO4X MRO5X ...
Страница 14: ...MRO2X 5X Manual 10 98 0058 Rev R 3 4FLUID DIAGRAM MRO2X MRO3X MRO4X ...
Страница 15: ...MRO2X 5X Manual 11 98 0058 Rev R 3 5 FLUID DIAGRAM MRO5X ...
Страница 35: ...MRO2X 5X Manual 31 98 0058 Rev R FIGURE 6 2 ...
Страница 36: ...MRO2X 5X Manual 32 98 0058 Rev R FIGURE 6 3 ...
Страница 52: ...MRO2X 5X Manual 48 98 0058 Rev R SECTION 11 SPARE PARTS LISTING ...
Страница 53: ...MRO2X 5X Manual 49 98 0058 Rev R ...