3
4.
Install the third long hose into the outlet of
the pump (the outlet is the port facing
upward on the pump).
5.
Install the short hose (approximately 2’ long)
into the pump inlet (the port at the end of the
pump that is facing horizontal).
6.
Connect the pump inlet hose to the fitting at
the bottom of the CIP tank using the cam-
lock connector.
7.
Turn off the MRO before connecting the CIP2 system or breaking any
water lines.
8.
Choose the correct fittings supplied with the CIP2 system that will mate to
the INLET, REJECT, and PRODUCT ports of the reverse osmosis system
to be cleaned, and install them.
9.
Connect the CIP pump outlet hose to the reverse osmosis water INLET
only.
WARNING: If the PRODUCT hose (PERMEATE) is momentarily pressurized due
to an incorrect connection (the CIP in reverse flow), the membranes
will be destroyed.
NOTE:
When connecting the CIP2 to an AmeriWater MRO 3, 4 or 5, the existing
MRO hoses (one ¾” hose and two ½” hoses) will need to be cut about 12” from
the hose ends opposite the MRO. Install the Cam-Lock male/female connector
pair using the included hose barbs and hose clamps. Be sure that the Cam-Lock
connectors are “gendered” in such a manner that the hoses can be connected to
the correct fittings on the CIP2 tank (see below).
10.
Connect the REJECT AND PRODUCT hoses to the CIP hoses near the
top of the CIP tank (hoses may be connected in either orientation).
11.
Verify that all connections are made and are watertight.