background image

TABLE OF CONTENTS 

1.

OBJECTIVE ................................................................................................... 1

2.

CHEMICALS REQUIRED .............................................................................. 1

3.

SET UP .......................................................................................................... 2

4.

INSTALLATION DIAGRAM ........................................................................... 4

5.

HIGH pH CLEANING ..................................................................................... 6

6.

LOW pH CLEANING ..................................................................................... 8

7.

AMERIWATER IRO SPECIFIC INSTRUCTIONS ........................................ 10

8.

SPARE PARTS LIST ................................................................................... 10

Содержание CLEAN IN PLACE CIP2

Страница 1: ...Part Number 98 0008 Revision G CLEAN IN PLACE MODEL CIP2 Manufactured With Pride In The USA www ameriwater com 800 535 5585 AmeriWater 3345 Stop 8 Rd Dayton OH 45414...

Страница 2: ......

Страница 3: ...TABLE OF CONTENTS 1 OBJECTIVE 1 2 CHEMICALS REQUIRED 1 3 SET UP 2 4 INSTALLATION DIAGRAM 4 5 HIGH pH CLEANING 6 6 LOW pH CLEANING 8 7 AMERIWATER IRO SPECIFIC INSTRUCTIONS 10 8 SPARE PARTS LIST 10...

Страница 4: ...e foulants from polyamide thin film composite 3 lb 1 110 ml of AmeriClean B powder is required per membrane cleaning AmeriClean B effectively removes the following foulants Silt Microbiological Colloi...

Страница 5: ...lean water to remove any debris or contamination after un packing or after the CIP system has been in storage 3 Install two of the long hoses into the upper ports on the tank by screwing the 1 pipe th...

Страница 6: ...ump outlet hose to the reverse osmosis water INLET only WARNING If the PRODUCT hose PERMEATE is momentarily pressurized due to an incorrect connection the CIP in reverse flow the membranes will be des...

Страница 7: ...ided to fit the AmeriWater MRO3 4 5 MROX MROZ models AmeriWater Industrial RO systems and other common RO systems Select the correct fittings fittings included to mate to the RO system at your facilit...

Страница 8: ...THE QUICK CONNECT FITTING CONNECT CONCENTRATE OUTLET TO THE CONCENTRTAE PORT ON THE CIP TANK DISCONNECT THE WEDGIT QUICK CONNECT FROM THE PERMEATE OUTLET CONNECT THE PERMEATE OUTLET TO THE PERMEATE PO...

Страница 9: ...ING Wear appropriate eye and skin protection when handling the membrane cleaning chemicals The cleaners are high and low pH chemicals and may cause severe chemical burns Read the MSDS for further info...

Страница 10: ...ations prior to emptying the cleaning solution into the drain Use an approved waste container for chemical waste disposal if required 13 After removing the spent chemical from the CIP2 tank and with t...

Страница 11: ...eaning solution is recirculating and flowing from the drain hose inside the CIP tank 6 Allow the cleaning solution to recirculate for a minimum of 50 minutes a Depending on the nature of the fouling a...

Страница 12: ...normal fault alarms to be active 13 Close the cover on the MRO and turn on the electrical power 14 To return the MRO to service close the PRODUCT AND REJECT valves going to the CIP2 then open the IN S...

Страница 13: ...ut of the normal discharge hoses for PRODUCT WATER or REJECT WATER 2 Turn OFF the power disconnect prior to opening the controller 3 After placing the CIP Switch in the ON position close the controlle...

Страница 14: ...INSx1IN FPT POLYGLASS 16 14520215 Adapter 1 0 Male NPT X 1 0 HB PP 17 15650100 CLAMP HOSE 1 SST 18 041531829 NIPPLE 75 X CL PVC80 19 046530953 Bush 1 0 X 0 75 THRD PP 20 16UMA100 Q CON COUP THD STEM x...