background image

OPERATION MANUAL FOR HEATSAN   

98-0177A 

  Page 2 

3 PREPARATION 

Prepare to upgrade the existing MRO by first removing the device from power.  
This needs to be done for the power to the control as well as the power to the 
pump.  Once the power has been removed, it is recommended to lock this out 
until the upgrade is complete. 

Loosen the 2 retaining screws on the front of the control panel to gain access 
to the CPU. 

Gently pry the CPU out of the socket, being careful not to break any of the 
pins.  Discard the old CPU. 

The empty socket is shown above.  Note the orientation of the notch at the 
bottom of the socket.  The replacement chip will have a notch to ensure that 
the CPU is keyed properly to allow proper operation

.  FAILURE TO ORIENT 

THESE IN THE SAME DIRECTION WILL LEAD TO DAMAGE TO THE 
CONTROLLER. 

Содержание 98-0177A

Страница 1: ...OPERATION MANUAL FOR MRO CONTROLLER UPGRADE TO HEATSAN 98 0177A UPGRADE AN MRO FOR USE WITH HEATSAN 98 0177A 02 16 2015...

Страница 2: ...OPERATING MANUAL FOR HEATSAN 98 0177A...

Страница 3: ...OPERATION MANUAL FOR HEATSAN 98 0177A Page 3 CONTENTS PAGE 1 DOCUMENT HISTORY 1 2 CONTACT DETAILS 1 3 PREPARATION 2 4 CONTROLLER OPERATION 4 5 CONTROLLER STANDARD SETPOINTS 8...

Страница 4: ...ription of Change A 17 02 15 Initial Release 2 CONTACT DETAILS For all service spares and consumables enquiries contact AmeriWater 1303 Stanley Ave Dayton Ohio 45404 Tel No 800 535 5585 Or your local...

Страница 5: ...upgrade is complete Loosen the 2 retaining screws on the front of the control panel to gain access to the CPU Gently pry the CPU out of the socket being careful not to break any of the pins Discard t...

Страница 6: ...o place Ensure that the chip is fully seated in the socket Return all power to the device Move the write protect jumper to the OFF position Power on the RO and access the program screen labelled as DI...

Страница 7: ...mp starts OPERATING The RO unit is operating TANK FULL The unit is shut down due to a tank full condition TANK FULL 99 The unit is shut down due to a tank full condition If the number is blinking the...

Страница 8: ...the switch is connected to the tank full high input When this switch has been active for 5 seconds the unit will shut down on tank full TANK FULL will show on the display When the tank full condition...

Страница 9: ...s active for the delay programmed in the PF Delay Setpoint the unit will shut down for a pressure fault The display will show PRESS FAULT the alarm lamp will flash and the audible alarm will sound The...

Страница 10: ...is warning will clear when the Conductivity drops below the Setpoint When the High Conductivity warning message is active the MRO will divert the PRODUCT WATER to drain through the Reject hose until t...

Страница 11: ...ty in either S or PPM 0 S 1 PPM 0 1 0 Switch Select Selects if the switch inputs are normally open or normally closed 0 32 0 Temp UOM Selects display of the temperature in either F or C 0 F 1 C 0 1 0...

Страница 12: ...inutes 3 Auto Reset When a pressure fault shut down is active the system will attempt to restart after this delay If set to 0 the system must be manually reset 0 99 minutes 0 Press Fault Delay The tim...

Отзывы: