background image

OPERATING MANUAL FOR HEATSAN   

98-0177A 

  Page 7 

Pretreat 

If the pretreat input becomes active (closes) and stays active for 2 seconds, 
the unit will shut down in a pretreat lockout condition. PRETREAT will show 
on the display and the unit will remain shut down as long as the pretreat input 
is active. 

High Conductivity 

If the Conductivity reading exceeds the limit programmed the Cond Limit 
Setpoint for the delay programmed in the Cond Delay Setpoint, the alarm 
lamp will light and the HI COND warning message will show on the display. 
This warning will clear when the Conductivity drops below the Setpoint.  

When the High Conductivity warning message is active, the MRO will divert 
the PRODUCT WATER to drain (through the Reject hose), until the Product 
water conductivity goes back into the acceptable quality range. 

Direct Feed 

This is the setting that installation of this new CPU enables.  Setting this to “1” 
allows for pairing with the Heatsan disinfection system.  This changes the 
operation of the “tank full high” and “tank full low” inputs.  “Tank full high” is 
used to indicate that the Heatsan is in operation while “tank full low” is used to 
turn the RO on and off when water is required.  Changing this set point to “0” 
allows for the RO to operate as if the CPU was not installed.  

Содержание 98-0177A

Страница 1: ...OPERATION MANUAL FOR MRO CONTROLLER UPGRADE TO HEATSAN 98 0177A UPGRADE AN MRO FOR USE WITH HEATSAN 98 0177A 02 16 2015...

Страница 2: ...OPERATING MANUAL FOR HEATSAN 98 0177A...

Страница 3: ...OPERATION MANUAL FOR HEATSAN 98 0177A Page 3 CONTENTS PAGE 1 DOCUMENT HISTORY 1 2 CONTACT DETAILS 1 3 PREPARATION 2 4 CONTROLLER OPERATION 4 5 CONTROLLER STANDARD SETPOINTS 8...

Страница 4: ...ription of Change A 17 02 15 Initial Release 2 CONTACT DETAILS For all service spares and consumables enquiries contact AmeriWater 1303 Stanley Ave Dayton Ohio 45404 Tel No 800 535 5585 Or your local...

Страница 5: ...upgrade is complete Loosen the 2 retaining screws on the front of the control panel to gain access to the CPU Gently pry the CPU out of the socket being careful not to break any of the pins Discard t...

Страница 6: ...o place Ensure that the chip is fully seated in the socket Return all power to the device Move the write protect jumper to the OFF position Power on the RO and access the program screen labelled as DI...

Страница 7: ...mp starts OPERATING The RO unit is operating TANK FULL The unit is shut down due to a tank full condition TANK FULL 99 The unit is shut down due to a tank full condition If the number is blinking the...

Страница 8: ...the switch is connected to the tank full high input When this switch has been active for 5 seconds the unit will shut down on tank full TANK FULL will show on the display When the tank full condition...

Страница 9: ...s active for the delay programmed in the PF Delay Setpoint the unit will shut down for a pressure fault The display will show PRESS FAULT the alarm lamp will flash and the audible alarm will sound The...

Страница 10: ...is warning will clear when the Conductivity drops below the Setpoint When the High Conductivity warning message is active the MRO will divert the PRODUCT WATER to drain through the Reject hose until t...

Страница 11: ...ty in either S or PPM 0 S 1 PPM 0 1 0 Switch Select Selects if the switch inputs are normally open or normally closed 0 32 0 Temp UOM Selects display of the temperature in either F or C 0 F 1 C 0 1 0...

Страница 12: ...inutes 3 Auto Reset When a pressure fault shut down is active the system will attempt to restart after this delay If set to 0 the system must be manually reset 0 99 minutes 0 Press Fault Delay The tim...

Отзывы: