24
98-0157 Rev P
5.4
Commissioning
ONLY BY CERTIFIED AMERIWATER REPRESENTATIVES.
WARNING:
This Reverse Osmosis System (RO) contains a preservative solution to prevent
microbiological growth and freezing. Discard all product water for at least two hours of
operation before placing the RO in service.
5.4.1
Pretreatment Lockout
The
MediQA
can utilize pretreatment lockout. Pretreatment lockout will prevent the device from
operating while the pretreatment is going through regeneration. The MediQA is shipped from the
factory with a jumper installed between wire terminal X24 and wire terminal 1. To enable the
pretreatment lockout functionality, remove the jumper wire between wire terminal X24 and wire
terminal 1. After removing the jumper wire, wire the normally closed side of the pretreatment
lockout switch or switches (located in the pretreatment controller) as shown below.
For Clack Valves:
Содержание 00MDP2-230
Страница 2: ......
Страница 61: ...59 98 0157 Rev P 9 2 1 MSP Spare Parts ...
Страница 62: ...60 98 0157 Rev P ...
Страница 63: ...61 98 0157 Rev P 9 2 2 MDP Spare Parts ...
Страница 64: ...62 98 0157 Rev P ...
Страница 66: ...64 98 0157 Rev P FLOW DIAGRAMS MSP UNITS ...
Страница 67: ...65 98 0157 Rev P MDP UNITS ...
Страница 68: ...66 98 0157 Rev P WIRING SCHEMATIC MSP UNITS ...
Страница 69: ...67 98 0157 Rev P ...
Страница 70: ...68 98 0157 Rev P ...
Страница 71: ...69 98 0157 Rev P ...
Страница 72: ...70 98 0157 Rev P ...
Страница 73: ...71 98 0157 Rev P ...
Страница 74: ...72 98 0157 Rev P ...
Страница 75: ...73 98 0157 Rev P ...
Страница 76: ...74 98 0157 Rev P MDP Units ...
Страница 77: ...75 98 0157 Rev P ...
Страница 78: ...76 98 0157 Rev P ...
Страница 79: ...77 98 0157 Rev P ...
Страница 80: ...78 98 0157 Rev P ...
Страница 81: ...79 98 0157 Rev P FX3U 4AD ADP ...
Страница 82: ...80 98 0157 Rev P ...
Страница 83: ...81 98 0157 Rev P ...
Страница 86: ...84 98 0157 Rev P ...
Страница 87: ...85 98 0157 Rev P MDP Units ...
Страница 88: ...86 98 0157 Rev P ...