34
EN
Replacing the fuses
WARNING
Risk of electric shock!
Disconnect the product from any circuit, remove
the connector leads (N) from the input jacks (F)/(G)/
(K)/(L), and turn OFF the product before opening
the battery compartment (R). Do not operate the
product with an open battery compartment.
NOTICE
If the product does not work
properly, check the fuses and batteries to make
sure that they are still good and that they are
properly inserted.
NOTICE
Always use a fuse of the proper size
and value:
– 800 mA/1000 V fast blow for the 600 mA
range,
– 10 A/1000 V fast blow for the 10 A range).
• Open the tilt stand (S).
• Loosen the screw of the battery compartment
cover (R) and remove it.
• Remove the old fuse and install the new fuse
into the holder.
• Close battery compartment (R) and tighten the
screw.