4

Arithmetic and Wind Speed/Direction mode functions

The LD display setup and calibration functions are configured through a push button sequence. The
three push buttons located on the input pcb (also at the front on some display options) are used to alter
settings. Two basic access modes are available:

FUNC

mode (simple push button sequence) allows access to commonly set up functions such as alarm

setpoints.

CAL

mode (power up sequence plus push button sequence) allows access to all functions including cali-

bration parameters.

Once

CAL

or

FUNC

mode has been entered you can step through the functions, by pressing and releasing

the

F

push button, until the required function is reached. Changes to functions are made by pressing the

^

or

v

push button (in some cases both simultaneously) when the required function is reached.

F

F

F

F

F

Entering

Mode

CAL

Entering

Mode

FUNC

1. Remove power from the instrument

and wait 5 seconds . Hold in the

button and reapply power.

The display will indicate

as part of the

"wake up messages" when

the          message is seen

you can release the

button.

Move to step 2 below.

2. When the "wake up"

messages have finished

and the display has settled

down to its normal reading

press, then release the

button.

Move to step 3 below.

1. When the "wake up"

messages have finished

and the display has settled

down to its normal reading

press, then release the

button.

Move to step 2 below.

3. Within 2 seconds of

releasing the      button

press, then release

the      and      buttons

together. The display will

now indicate             followed

by the first function.

2. Within 2 seconds of

releasing the      button

press, then release

the      and      buttons

together. The display will

now indicate             followed

by the first function.

^

^

^

^

FUNC

FUNC

F

F

F

F

F

CAL

CAL

Notes: If step 1 above has been completed then the

instrument will remain in this

mode state until

power is removed. i.e. there is no need to repeat step 1

when accessing function unless power has been removed.

The instrument should show all 8’s on power up

e.g.

if the instrument does not reset then

these numbers will not be seen. Switch off the instrument

and allow a longer time delay before powering up again.

CAL

8.8.8.8.

No special power up procedure
is required to enter

mode.

FUNC

Alternative

Mode Entry

CAL

This alternative method allows

mode

entry without the need to remove power:

Enter

mode using the 2 steps above

When the first function appears press and

hold the

button until you see the message

followed by

(the

button will have

to be held pressed for approximately 2 seconds)

You should now return to the function you were

in but have full access to

mode functions

Note: when you exit back to live reading the display
will remain in

mode for approximately 4

minutes, after this time you will need to repeat
this process to enter

mode.

CAL

FUNC

FUNC

CAL

CAL

CAL

CAL

1.

2.

3.

P

P

F

F

F

Easy alarm relay adjustment access facility

The display has an easy alarm access facility which allows access to the alarm setpoints simply by pressing
the

F

button at the front or rear of the instrument. The first setpoint will then appear and changes to

this setpoint may be made to this setpoint via the

^

or

v

buttons. Press the

F

button to accept any

LDRSMANADD-2.3-0

15 of

44

Содержание LC-CL Series

Страница 1: ...lay Arithmetic and Wind Speed Direction Addendum AMALGAMATED INSTRUMENT CO PTY LTD ACN 001 589 439 Unit 5 28 Leighton Place Hornsby Telephone 61 2 9476 2244 e mail sales aicpl com au NSW 2077 Australi...

Страница 2: ...allation 4 3 Function tables Arithmetic mode 10 4 Arithmetic and Wind Speed Direction mode functions 15 5 Arithmetic function examples 40 6 Function table Wind Speed and Direction mode 41 7 Wind speed...

Страница 3: ...o 20 0 seconds The instruments polled for arithmetic operation must be of the same manufacture as this instrument To operate in this mode the COdE function must be set to ARth Wind speed and direction...

Страница 4: ...is manual to confirm proper selection of voltage polarity and input type before applying power to the instrument When power is applied the instrument will cycle through a display sequence indicating t...

Страница 5: ...olated DC supply E IN Optional isolated dc board for use with 100mm 200mm led displays and 100mm electromagnetic displays only CASE EARTH Input DC 12 24V Output DC to display board 2 5 100mm and 200mm...

Страница 6: ...ge instruments use a different transformer Do not use a low voltage transformer for 240V or 110V or vice versa 110VAC 42VAC 32 TO 36VAC 24VAC ACTIVE ACTIVE ACTIVE ACTIVE ACTIVE ACTIVE NEUTRAL NEUTRAL...

Страница 7: ...g RS232 to a 9 or 25 pin D type connector the standard connections are as follows 9 pin Tx pin 3 Rx pin 2 GND pin 5 25 pin Tx pin 2 Rx pin 3 GND pin 7 See note below regarding Tx and Rx crossing when...

Страница 8: ...ith connections on terminal P2 The relays are single pole single throw types and are rated at 5A 240VAC into a resistive load The relay contacts are voltage free and may be programmed for normally ope...

Страница 9: ...S MMW 005 wind speed and direction sensor The WS MMW 005 can be connected as either RS232 or RS485 not serial current loop use the connection type which matches the input type of the display LDRSMANAD...

Страница 10: ...e delay for alarm relay 1 0 0 to 999 9 0 0 4 9 19 A2rt Reset time delay for alarm relay 2 0 0 to 999 9 0 0 4 10 19 A1n o or A1n c Alarm relay 1 action to normally open or normally closed A1n o or A1n...

Страница 11: ...to t8 P1 4 31 24 Ch7 Addr Channel 7 address P1 to P8 or S1 to S8 or t1 to t8 P1 4 32 24 Ch8 Addr Channel 8 address P1 to P8 or S1 to S8 or t1 to t8 P1 4 33 24 Ch1 dCPt Channel 1 decimal point 0 0 1 et...

Страница 12: ...32767 1 4 53 28 EA5 EA5 value for arithmetic formula 32727 to 32767 1 4 54 28 Eb5 Eb5 value for arithmetic formula 32727 to 32767 0 4 55 28 EC5 EC5 value for arithmetic formula 32727 to 32767 1 4 56 2...

Страница 13: ...70 32 OP6 Arithmetic operation between channel 7 and previous operation Add Sub Prod diU High Lo SINE or COS Add 4 71 32 OP7 Arithmetic operation between channel 8 and previous operation Add Sub Prod...

Страница 14: ...ss mode OFF EASY NONE or ALL OFF 4 85 36 SPAC Setpoint access mode Optional A1 A1 2 etc A1 4 86 36 A1 Alarm relay 1 operation channel ch0 to ch8 ch0 4 91 38 A2 Alarm relay 2 operation channel ch0 to c...

Страница 15: ...now indicate followed by the first function 2 Within 2 seconds of releasing the button press then release the and buttons together The display will now indicate followed by the first function FUNC FU...

Страница 16: ...used then the only way to view or alter any other function settings is to power up via CAL mode i e there is no entry to FUNC mode functions unless the instrument is powered up in CAL mode 4 1 Alarm r...

Страница 17: ...high setpoint then the relay will activate when the value displayed moves outside the band set by the low and high setpoints The value at which the relay will reset is controlled by the A1HY function...

Страница 18: ...isplay value falls to 20 0 or below and will reset when the display value goes above 30 0 i e at 30 1 or above The hysteresis units are expressed in displayed engineering units Example If A1Hi is set...

Страница 19: ...A1rt is set to 10 0 seconds then the resetting of alarm relay 1 will be delayed by 10 seconds 4 10 Alarm relay 2 reset time Display A2rt Range 0 0 to 999 9 Default Value 0 0 Displays and sets the alar...

Страница 20: ...o 50 then Alarm 1 will activate at 1000 and alarm 2 will activate at 1050 i e 1000 50 If Alarm 2 had been set at 50 then alarm 2 would activate at 950 i e 1000 50 4 14 Display brightness Display brGt...

Страница 21: ...The high brightness level sets the maximum brightness which the automatic brightness control can achieve with 63 being the highest intensity 4 18 Automatic display brightness adjustment low level Dis...

Страница 22: ...ay may take to reach its final value when the input is changed similarly the relay operation and any output options will be slowed down when the filter setting is increased To set the digital filter v...

Страница 23: ...ndary values are available for that instrument The primary value is the main display value for that instrument e g For a conductivity instrument the conductivity would be the primary value and the tem...

Страница 24: ...26 for further information 4 32 Channel 7 address Display Ch7 Addr Range P1 to P8 or S1 to S8 or t1 to t8 Default Value P1 Seen only when Code function is set to SCAN or Arth Scan address for channel...

Страница 25: ...s and sets the decimal point for input channel 2 See function 4 34 for further details 4 38 Channel 5 decimal point Display Ch5 dCPt Range 0 0 1 etc Default Value 0 Seen only in CodE Arth mode Display...

Страница 26: ...Ch2 Eb2 EC2 Add Sub Prod diU hIGh Lo SinE Cos C SUb EA3 Ch3 Eb3 EC3 etc for the remaining active channels Note A and C are whole numbers b has the same decimal place setting as its associated channel...

Страница 27: ...isplays and sets the EA3 value to be used in the arithmetic formula See function 4 42 for further details 4 49 Channel 3 Eb3 value for arithmetic formula Display Eb3 Range 32727 to 32767 Default Value...

Страница 28: ...e EA5 value to be used in the arithmetic formula See function 4 42 for further details 4 55 Channel 5 Eb5 value for arithmetic formula Display Eb5 Range 32727 to 32767 Default Value 0 Seen only in Cod...

Страница 29: ...e EA7 value to be used in the arithmetic formula See function 4 42 for further details 4 61 Channel 7 Eb7 value for arithmetic formula Display Eb7 Range 32727 to 32767 Default Value 0 Seen only in Cod...

Страница 30: ...formula Sub channel 1 formula minus channel 2 formula Prod channel 1 formula times channel 2 formula diU channel 1 formula divided by channel 2 formula High highest of channel 1 formula or channel 2...

Страница 31: ...and OP2 times channel 4 formula diU mathematical result of OP1 and OP2 divided by channel 4 formula High highest of mathematical result of OP1 and OP2 or channel 4 formula Lo lowest of mathematical r...

Страница 32: ...formula Lo lowest of mathematical result of OP1 and OP2 and OP3 and OP4 or channel 6 formula Sine mathematical result of OP1 and OP2 and OP3 and OP4 times the sine of the angle represented by channel...

Страница 33: ...nd OP6 or channel 8 formula Lo lowest of mathematical result of OP1 and OP2 and OP3 and OP4 and OP5 and OP6 or channel 8 formula Sine mathematical result of OP1 and OP2 and OP3 and OP4 and OP5 and OP6...

Страница 34: ...and sets the polarity selection for the display of the engineering value for channel 3 See function 4 74 for further information 4 77 Channel 4 polarity Display Ch4 Range both POS or NEG Default Value...

Страница 35: ...e new mode ARTH mode Arithmetic mode In this mode the display can accept inputs from up to 8 external sources and combine them arithmetically using a flexible formula see the EA1 formula for details o...

Страница 36: ...to gain access to alarm and calibration functions If set to ALL then access to all functions including calibration functions can be gained via FUNC mode 4 86 Setpoint access mode Display SPAC Range A1...

Страница 37: ...function is set to UAL Remote input function When these remote input terminals are short circuited via a switch relay keyswitch etc the instrument will perform the selected remote input function A me...

Страница 38: ...brightness control The remote input can be used to change the display brightness When this mode is selected the display brightness can be switched via the remote input terminals between the brightnes...

Страница 39: ...for 8 data bits plus 1 stop bit or 7 bit for 7 data bits plus 1 stop bit This must be set to match the number of data bits of the sending device 4 97 Serial communications type Display SER 1 tYPE Rang...

Страница 40: ...est of the 3 inputs Channel 1 input is to be multiplied 2 3 Program EA1 2 Eb1 0 EC1 3 EA2 1 Eb2 0 EC2 1 EA3 1 Eb3 0 EC3 1 OP1 HiGH OP2 HiGH 2 Ch1 0 3 HIGH 1 Ch2 0 1 HIGH 1 Ch3 0 1 Example 3 To multipl...

Страница 41: ...9 0 0 4 8 19 A1rt Reset time delay for alarm relay 1 0 0 to 999 9 0 0 4 9 19 A2rt Reset time delay for alarm relay 2 0 0 to 999 9 0 0 4 10 19 A1n o or A1n c Alarm relay 1 action to normally open or n...

Страница 42: ...input external input one function NONE P HLd d HLd Hi Lo HiLo tARE ZERO SP Ac No Ac or duLL NONE 4 90 37 ACCS Access mode OFF EASY NONE or ALL OFF 4 85 36 SPAC Setpoint access mode Optional A1 A1 2 e...

Страница 43: ...t to 4800 with no parity PrtY function and 8 data bits dAtA function The display can be set to show wind speed wind direction or to toggle using or v button or automatically alternate the display betw...

Страница 44: ...SO N 14 0E 9 57 39 d 100 64 SI O 15 0F 58 3A e 101 65 DLE P 16 10 59 3B f 102 66 DC1 Q 17 11 60 3C g 103 67 DC2 R 18 12 61 3D h 104 68 DC3 S 19 13 62 3E i 105 69 DC4 T 20 14 63 3F j 106 6A NAK U 21 15...

Отзывы: