ALT I076 Issue 03/0418
Page 23
4.5 Operation: Refilling bottle
STEP 1
Slide back the quick disconnect single connector (
Item 3
) to separate the green air supply tubing
(
Item 6
) from the bottle cap (
Item 1
). This is done first to release any remaining pressure in the
system and to avoid spilling excess fluid.
STEP 2
Depress the silver plate on the quick disconnect dual connector (
Item 2
) to separate the air/fluid
tubing (
Item 5
) from the bottle cap (
Item 1
)
STEP 3
Take the bottle from well and remove the cap by turning counter-clockwise.
STEP 4
Fill the bottle with the sterile water for irrigation (or similar), leaving one inch of space from the
top of the bottle to ensure proper operation.
STEP 5
Replace the bottle cap. Place bottle back in the well and reconnect the green air supply tubing
(
Item 6
) and air/fluid tubing (
Item 5
). Ensure both quick disconnect connectors are secure.