4.2 Operation: Automated Cycle
STEP 1
See Caution #1 and complete all pre-use checks as advised in Table 1 (see page 14 above).
Connect the instrument interface tube (
Item 17
) to the item to be rinsed (Figure 6), use the
adaptor pack (
Item 15
) and/or
A11047 “Y” Dual Rinse Fitting
if necessary.
STEP 2
Press the Cycle button (
Item 12
) once to begin the automated rinse cycle. The amber light next
to the fluid valve will light up indicating that the valve is open. After 15 seconds the fluid valve
will close and the air valve open. The amber light next to the air valve will now be illuminated.
After 15 seconds the air valve will close, the machine will stop and both amber lights will be
turned off. This completes the automated cycle. See Table 2 (page 26 below) for Troubleshooting
if necessary.
STEP 3
Repeat Steps 1 and 2 for any further lumen on the device.
STEP 4
Inspect the rinsed device for cleanliness and function prior to further processing, e.g. disinfection
or sterilization.
* See Page 8 for “Item” References