9 Service and Maintenance
9.1 General Information
NOTICE
Damage by insertion of objects
Inserting any objects into the measurement aperture could damage the in-
strument. Do not insert any objects into the measurement aperture.
NOTICE
Damage by using acetone for cleaning
Instrument housing, white and green standard tile can be damaged when
they get in contact with acetone. Do not use acetone for cleaning.
NOTICE
Damage by the attempt of self-repair
The instrument can be damaged. Warranty claims expire. Do not attempt
to make any repairs yourself. Contact our customer service in case of mal-
function.
9.2 Storing the Instrument
Align the protective cap when the instrument is not in use. Use the instru-
ment case for storage.
9.3 Cleaning the Instrument
Use a soft, moist cloth for cleaning. For cleaning excessive dirt, use
propanol.
9.4 Cleaning the Standard Tiles
Using dirty or damaged standard tiles can impact the accuracy of measure-
ments significantly. Clean the standard tiles when they are dirty in order to
avoid measuring inaccuracy.
NOTICE! Apply only slight pressure as you clean and make certain
there are no large particles stuck in the cloth that could damage
the surface. We highly recommend to handle the standard tiles
with great care. Store them always enclosed.
Use a new lint-free cloth, dust-free lens paper or an optical cloth for clean-
ing. For dirt that is difficult to remove, use an optical cloth dipped in
propanol. Then wipe the surface with a dry optical cloth.