6.3 Instrument Calibration
The system delivery comprises a calibration tile and test tiles for color and
gloss, see Test and Calibration Standards [
16]. These tiles can be used to
check if your instrument is measuring correctly. Each tile comes with a dedi-
cated serial number matching your instrument. In the configuration menu
select
Calibration
.
1
2
3
4
5
6
7
8
9
10
1
Information
Check date of last calibration,
serial numbers and values for
calibration and test tiles here.
6
White standard
Comprises the options for in-
strument calibration.
2
Autodiagnosis
Comprises the options for auto
diagnosis.
7
Calibrate on white stan-
dard
Start the manual calibration
procedure here.
3
Docking station interval
Select interval for the auto di-
agnosis here. Possible values:
1, 3, 7, 14 days. Default: 7
days.
8
Add standard to list
Add calibration tile to the list
of standards in the instrument.
4
Perform diagnosis
Start the auto diagnosis proce-
dure here.
9
Green standard
Comprises the options for in-
strument test.
5
Add docking station
Turn your spectro2go into
spectro2guide here or replace
existing docking station.
10
Add standard to list
Add test standard to the list of
standards in the instrument.
NOTES:
• The option
Add standard to list
is active, if the standard has not been
added yet.
• In order to delete these standards from memory use the function
37].